47
The alarm delay time delays the color change of the display backlighting to red, the
22 mA signal (if configured), and the alarm contact switching.
Error messages can be signaled by a 22 mA output current. In addition, a relay
contact (RELAY1 / RELAY2) can be configured as alarm contact.
Menu item
Action
Choices
Alarm delay
Enter value using
keys.
Press enter to confirm.
0...600 SEC
(10 SEC)
Sensocheck
Select Sensocheck (continuous
monitoring of sensor membrane
and lines).
Select ON or OFF using
keys.
Press enter to confirm.
(At the same time, Sensoface is
activated. With OFF, Sensoface is
also switched off.)
ON/OFF
Tempcheck
To monitor the temperature
probe with TC OFF selected:
Select Tempcheck ON using
keys.
Press enter to confirm.
Now, the temperature probe will
be monitored.
ON/OFF
Configuring the Alarm