CondI
Note:
When the sensor is installed in a
pipe/tank at a distance less than
30 mm from the wall, you should
perform the calibration either by
means of sampling (product cali-
bration) with the sensor installed or
in a suitable calibration beaker with
dimensions and material corre-
sponding to the process conditions.
Calibrating Toroidal Sensors