Operation
21
Diagnostics level
On the Diagnostics level the complete pH measure-
ment equipment is checked. This also serves for
quality management to ISO 9000.
Activating diagnostics
Press
to exit measuring mode and access
the main menu of the Diagnostics level. To exit
Diagnostics level, press
. It is exited auto-
matically when all diagnostics functions have been
performed.
Main menu
In the main menu you select the desired function of
the Diagnostics level using
or
. Pressing
gives access to the corresponding sub-
menu.
The Sensoface
®
menu shows the states of the cri-
teria that control Sensoface
®
display.
The Electrode Data menu shows the electrode data
calculated during last calibration, the calibration
timer count and the electrode statistics.
The Knick Fullcheck
®
menu provides a complete
device self test to check the operability of the indi-
vidual device components.
All submenus are run automatically but can be influ-
enced via keypad.
• To stop the sequence, press
. To restart it,
press
once more.
• To scroll one step forwards or backwards, press
or
, respectively.