765 Laboratory pH Meter
40
Operation
Only for meters
with Option 346:
Nominal electrode zero and slope are set as fol-
lows:
Control commands
Reading device description
Command
Parameter
Unit
Description
WPCA0NZ
Num. par.
[pH]
Nominal
zero point
WPCA0NS
Num. par.
[V/pH]
Nominal
slope
WPCA0U
Num. par.
[V]
V iso isothermal
potential
Command
Description
WCIU
Initialize device
WCTEA
Perform Fullcheck
WCRTT [hhmm]
Set time
WCRTD [ddmmyy]
Set date
WCOM00
Switch to measuring mode
WCCASTI
Set first calibration
(only for Option 347)
WCCAA1
Start calibration with 1st buffer
WCCAA2
Start calibration with 2nd buffer
WCDISLA0
Left display, indicate pH
WCDISLA1
Left display, indicate mV
WCDISLA2
Left display, indicate temperature
WCDISLATRT
Left display, indicate time
WCDISRA0
Right display, indicate pH
WCDISRA1
Right display, indicate mV
WCDISRA2
Right display, indicate
temperature
WCDISRATRT
Right display, indicate time
Com-
mand
Response
Description
RDMF
KNICK
Manufacturer
RDUN
765
Model name
RDUS
xxxxxx
Serial number
RDUV
xx;xx
Software/hardware version
RDUP
xxx;xxx;xxx
Options