Operation
13
Sensoface
®
was specially designed for monitoring
pH electrodes. Since ORP electrodes have com-
pletely different characteristics, they provoke
invalid Sensoface
®
displays. Therefore, you should
disable Sensoface
®
for ORP measurements. Sen-
soface
®
display can also be degraded when you
use a pH simulator.
Displaymatic
®
(off / on)
Displaymatic
®
can be turned on or off. Display-
matic
®
ensures a stable measured value display in
case of rapid changes of the pH value. This allows
for easy reading of the value. It is achieved by sup-
pressing the two running digits behind the decimal
point. When the value changes slowly, only the last
digit will be suppressed. When the value has stabi-
lized, all digits will be displayed.
Only for meters
with Option 346:
Setting the nominal electrode zero point and slope:
Nominal zero (pH 0.00 to 14.00)
For special electrodes with differing nominal zero.
The entered zero point is taken as reference for
Calimatic
®
. The exact zero point will be determined
during calibration.
Nominal slope (25.0 to 61.0 mV/pH)
For special electrodes with differing nominal slope.
The entered slope is taken as reference for Cali-
matic
®
. The exact slope will be determined during
calibration.
Isothermal potential
(–500 to +500 mV)
The isothermal potential entered here is automati-
cally taken into account for calibration and pH
determination.