Operation
5
Menu structure
Sensoface
®
automatic
electrode monitoring
Sensoface
®
automatic electrode monitoring pro-
vides information on the electrode condition. Zero
point, slope and response time are evaluated dur-
ing calibration.
Impedance, dry-out and the calibration timer are
evaluated during measurement and in standby
mode.
A summary of the individual results is expressed by
three face symbols.
The electrode is in good condition.
The electrode is still usable, but should be cali-
brated.
The electrode urgently requires calibration. If the
display does not change after calibration, replace
the electrode.