Operation
Vacuum Pump System SC/SEM/SH/SR950
28
Translation of original Operating Instructions, English, KNF 125549-125552 07/18
Only SC 950:
DANGER
Uncontrolled operation may result in personal injury
and damage to the vacuum pump system.
When using several vacuum pump systems SC 950,
there is the danger of confusing them, which can
result in undesired interference into other processes:
The transmission of commands from the wrong hand
terminal for the vacuum pump system may trigger an
uncontrolled response in the vacuum pump system
that the terminal controls.
Every time a vacuum pump system SC 950 is
used with a hand terminal, it is essential to
ensure a correct match between the two
components. Use the paging function for this
purpose (see Actuating the vacuum pump
system, page 38).
Additionally it is possible to use color stickers to
indicate which hand terminal belongs to which
vacuum pump system SC 950 (see chapter
Spare parts and accessories
).
Only SC 950:
WARNING
Uncontrolled operation may result in personal injury
and damage to the vacuum pump system SC 950.
If the wireless connection between the hand terminal
and vacuum pump system SC 950 is broken, the
vacuum pump system will continue to operate with
the current settings.
Immediately determine and remove the cause of
the interruption between the hand terminal and
vacuum pump system (chapter 10, page 54).
If you are not able to reestablish wireless
contact, replace the hand terminal back to the
vacuum pump system (chapter 8.1.1, page 31).
The vacuum pump system can also be operated
directly and the ventilation valve and coolant
valve can be opened and closed directly
(chapter 8.2, page 40).