Vacuum Pump System SC/SEM/SH/SR950
Use
Translation of original Operating Instructions, English, KNF 125549-125552 07/18
5
2. Use
2.1. Intended use
The vacuum pump systems SC/SEM/SH/SR 950 are designed for
use in chemical, pharmaceutical, and biological laboratories. They
are exclusively intended for transferring gases and vapors.
Make sure that the installation location is dry and the pump/system
is protected against water in the form of rain, spray, splashes and
drips.
Vacuum pump system is solely for use in indoor areas.
Owner's responsibility
Only install and operate the vacuum pump system under the
operating parameters and conditions described in chapter 4,
Technical data.
Protect vacuum pump system against humidity.
Before using a medium, check the compatibility of the materials of
the pump head, diaphragm, valves, sealings and tubing with the
medium.
Before using a medium, check whether the medium can be trans-
ferred danger-free in the specific application case.
Only transfer gases which remain stable under the pressures and
temperatures occurring in the pump.
Only SC/SEM/SH 950:
The high performance condenser must be installed on the outlet
side of the pump; if it is installed on the inlet side there is a danger
of implosion.
Observe the correct usage of the gas- and cooling liquid-
connections on the high performance condenser. Inlet and outlet
connections for the gas are not interchangeable.
Laboratory equipment or additional components connected to a
vacuum pump system have to be suitable for use with the pneu-
matic capabilities of the vacuum pump system (see chapter 4,
page 8).
Operating parameters and
conditions
Requirements for
transferred medium
High performance condenser
Accessories