10
AZURA® Valve Unifier VU 4.1 Instructions V6855
6. Unpacking and setup
This chapter describes all preparatory steps prior to start-up.
6.1
Operating environment
Only if the requirements for ambient conditions of the operating environ-
ment are met, can the intended use be ensured. You will find the ambient
condi tions under Technical Data.
Device defect
The device overheats at exposure to sunlight and insufficient air circula-
tion. Device failures are very likely.
Î
Set up the device in such a way that it is protected against exposure
to direct sunlight.
Î
Leave room for air circulation: See paragraph „space requirements“.
At least 5 cm, if there is another device on one side.
At least 10 cm, if there are devices set up on both sides.
Position the device on a level surface.
Protect the device against direct exposure to sunlight.
Set up the device at a location not exposed to air drafts (A/C systems).
Do not set up the device near other machines that cause floor
vibrations.
Avoid sources of high frequencies near the device. High-frequency
sources may compromise measuring values.
6.2
Unpacking the device
Check packaging for damage caused during transportation. If neces-
sary, put forward any claim for damages to the carrier.
Utility knife
Bruising danger
Damage to the device by carrying or lifting it on protruding housing
parts. The device may fall and thus cause injuries.
Î
Lift the device only centrally on the side of the housing.
1.
Set up the package in such a way that you can read the label. Using
the utility knife, cut the adhesive tape and open the packaging.
2.
Remove the foam insert. Take out the accessory kit and the manual.
3.
Open the accessory kit and check the scope of delivery. In case any
parts are missing, contact the Technical Support.
Space requirements
General requirements
Prerequisite
Tool
Process