SMY133-SMY134-SMP133 Operating Manual
4.3.2 Measured Values Evaluation and Aggregation
As described above, measured values are evaluated according to IEC 61000-4-30 ed.2, based on
continuous (gap-less), 10 / 12 mains cycles long intervals ( measurement cycle ) processing.
Further aggregation of the actual values from this evaluation is used to obtain values for displaying
and recording.
4.3.2.1 Actual Values Evaluation and Aggregation
Actual ( instantaneous ) values of measured quantities, that can be viewed on instrument's display,
are evaluated as average of integral number of measurement cycle values per
display refresh cycl
e.
The display refresh cycle is presetable in range 2 - 20 measurement cycles , corresponding approx. to
0.4 - 4 seconds display refresh cycle duration.
Maximum (marked as
↑
) and minimum (
↓
) measurement cycle values registered during the display
refresh cycle interval are displayed too.
Exceptions :
•
frequency – the value is refreshed each frequency measurement cycle (see above)
•
harmonic components, THD and unbalance – the last measurement cycle values are
displayed (no averaging)
•
temperature – the value is refreshed each temperature measurement cycle (see above)
Actual values, read from an instrument via a communication link for monitoring purposes are
evaluated from one – the last – measurement cycle only.
Neither maximum nor minimum of cosφ values are evaluated due to special character of the
quantity. Similarly, these extreme values are not evaluated at frequency, harmonics&THD
and temperature values either due to a specific measurement method.
4.3.2.2 Average Values Evaluation
From measurement cycle values, average values of all basic quantities are calculated. Following
parameters can be set to control the way of averaging :
•
averaging method, which can be set to one of :
•
fixed window
•
floating window
•
averaging period in range from 0.2 second to 1 hour
When
fixed window
averaging is set, values are calculated from fixed block intervals. The values are
updated at the end of every interval. Beginnings of the intervals are synchronized to the nearest whole
time ( for example, when averaging period is set to 15 minutes, the average values are updated four
times per hour in xx:00, xx:15, xx:30 and xx:45 ).
When
floating window
is set, the exponential moving average is evaluated.
28
Fig. 4.21 : Actual Data
Display Refresh Cycle
Setting
Fig. 4.22 : Actual Data
maximum measurement cycle
value during refresh cycle
minimum measurement cycle
value during refresh cycle
display refresh cycle
average value