![KLS Martin maXium smart C Скачать руководство пользователя страница 75](http://html1.mh-extra.com/html/kls-martin/maxium-smart-c/maxium-smart-c_instructions-for-use-manual_1979004075.webp)
maxium® smart C Electrosurgery Unit with maxium® smart Beam
Revision 1
75
Technical Data
Power setting range:
1–150 W at 800
Ω
Max. periodic HF output voltage:
2,000 V
p
at maximum effect setting
Fundamental frequency:
404 kHz
Crest factor at 600
Ω
:
3–5.2
Output curves
Output power versus set power
Repetitive peak output voltage versus set effect
HF
p
ow
er
in
w
at
ts
HF
v
ol
ta
ge i
n v
ol
ts
pe
ak
Setting
Effect
Output power versus resistance at 50% and 100% of max.
power setting
Repetitive peak output voltage versus set power
HF
p
ow
er
in
w
at
ts
HF
v
ol
ta
ge i
n v
ol
ts
pe
ak
Load resistance in
Ω
Setting