![KLS Martin maXium smart C Скачать руководство пользователя страница 125](http://html1.mh-extra.com/html/kls-martin/maxium-smart-c/maxium-smart-c_instructions-for-use-manual_1979004125.webp)
maxium® smart C Electrosurgery Unit with maxium® smart Beam
Revision 1
125
When the neutral electrode for babies is selected, only the following monopolar currents and maximum
powers are permitted:
Monopolar cutting currents
Monopolar coagulation currents
Pure Cut up to 30 W
Forced Clamp up to 30 W
Blend Cut up to 30 W
Forced Prep up to 30 W
Super Cut up to 30 W
Spray up to 30 W
Contact Coag up to 30 W
Forced Coag up to 30 W
Argon Beam up to 30 W
The bipolar current types are fully available.
In case of selection of the Mega Soft neutral electrode, only the following monopolar current types are
allowed:
Monopolar cutting currents
Monopolar coagulation currents
Blend Cut
Forced Clamp
Uro Blend Cut
Forced Prep
Spray
Contact Coag
Forced Coag
Uro Coag
Agon Beam
The bipolar current types are fully available.