Messages of the Unit
The bottom line of the screen text is used to display texts indicating operating or error conditions. Below,
the messages are explained which help the user to determine the current status of the unit or perform to
troubleshooting. All other messages are described in the maxium® Service Manual.
13.1
Messages during Self-Test
After switching on, the unit displays the startup screen. If the unit remains in the starting phase and any of
the following messages appears, then proceed as described.
Message
Description
Self-test: “Cut” foot switch engaged upon
startup!
Self-test: “Coag” foot switch engaged upon
startup!
Self-test: Universal foot switch engaged
upon startup!
Self-test: Backup foot switch engaged upon
startup!
The unit expects that during the self-test no activation element is
engaged. If during self-test is a foot switch is actuated, the unit will
notice this.
Check whether the cable is wrapped around the foot switch so that
the pedal is actuated.
Turn the unit off and then on again. If the message appears again,
disconnect the foot switch from the unit and turn the latter off and
on again. If the self-test is now completed successfully, the circuit of
the instrument is shorted, otherwise there is a malfunction.
Self-test: “Cut” finger switch before startup!
(Cable?)
Self-test: “Coag” finger switch before
startup! (Cable?)
The unit expects that during the self-test no activation element is
engaged. If during the self-test is a finger switch is actuated, the unit
will notice this.
Turn the unit off and then on again. If the message appears again,
disconnect the instrument from the unit and turn the latter off and
on again. If the self-test is now completed successfully, the circuit of
the instrument is shorted, otherwise there is a malfunction.
Self-test: Pressed or stuck key during system
start!
The unit expects that during the self-test no activation element is
engaged, otherwise the self-test will end with an error.
Turn the unit off and then on again. If the self-test is again not
completed successfully, there is a malfunction.
Self-test: Rotary switch active during system
start!
The unit expects that during the self-test no control element is
engaged. The unit detects the movement of the rotary switch during
the self-test.
Turn the unit off and then on again. If the self-test is again not
completed successfully, there is a malfunction.