maxium® smart C Electrosurgery Unit with maxium® smart Beam
44
Revision 1
•
To unmount the maxium® smart C and maxium® smart
Beam, insert a round object into the opening provided for
this purpose and simultaneously lift the device.
Perform the gas installation as follows:
•
Set up the pressure vessel with the argon gas so that it is secured against toppling.
•
Remove the protective cap and screw on the pressure reducer. Tighten the knurled screw of the
pressure reducer only manually!
•
Snap the hose of the pressure reducer onto the argon inlet at the rear of the beamer. If the angle valve
of the pressure vessel is already open, then a bit of force is required, and a small amount of gas will
escape in a loud hiss.
After turning the angle valve open by at least half a turn in counterclockwise direction, the system is ready
for commissioning.
Risk of injury by use of improper pressure reducer!
Do not connect any pressure reducer to the maxium® smart Beam other than that offered as a standard
accessory by Gebrüder Martin!