maxium® smart C Electrosurgery Unit with maxium® smart Beam
114
Revision 1
14
Periodic Safety Checks (SC)
At least every 24 months the system (maxium® smart C with maxium® smart Beam) must be subjected to
the following controls to be performed by people who, by virtue of their training, knowledge and expe-
rience gained from practical activities are able to properly perform such safety checks and not obliged to
adhere to instructions with regard to such control activities. If the maxium® smart C is equipped with a
maxium® smart Beam, this shall be inspected during the control of the maxium® smart C, too. A suitable
protocol form can be requested from the Martin Service Center.
A: Visual inspections and function checks
•
Readability of type plate and markings.
The information on the type plates of the maxium® smart C and, if applicable, maxium® smart Beam,
as well as the labels on the rear side of the units, in particular those indicating the nominal value for
the fuses of the units in the power cable connection, must be clearly legible. On the maxium® smart
Beam, the indication of the permissible pressure range at the gas inlet must be clearly legible.
•
The instructions for use must be available in a version suitable for the unit. Updated instructions for
use can be obtained from the Martin Service Center.
•
The software version number VX.xxx shown on the startup screen upon switching on of the unit is to
be entered into the protocol sheet. Note messages during the self-test.
•
Make sure the two activation lamps are working (uniform illumination of the field of view). Clear
rendering of the icons and correct reproduction of the colors blue and yellow must be verified.
•
Visual inspection of all ports on the unit.
–
Check the HF sockets M and B for visible damage.
–
Check sockets for foot switches, equipotential grounding connection and socket for maxium®
smart Beam on the rear for damage and tight fit.
–
Pull the mains cable out of the socket on the unit and test the connection socket for damage and
tight fit.
–
On the maxium® smart Beam, check connection socket for the connection cable to the maxium®
smart C for damage and tight fit.
–
Check gas inlet, decompression pin and grounding connection for tight fit.
–
Check the gas outlet on the front for damage and tight fit.
•
Check the cables of all reusable accessories used with the unit for damage to the insulation and the
connector plug, especially in the area of the particularly strained ends.
•
Function check of activation, activation lamps and activation sound.
–
Disable NE alarm with jumper plug,
–
connect electrode handle to M and
–
press the two finger switches.