Klinger CMF
Instruction Manual
Page 42
5.2.6 ADVANCED
5.2.6.1 In the “ADVANCED” of the secondary interface, press “FUNC” to enter the third level
interface. Set password: 995.
5.2.6.2 After entering third level interface, press “UP” or “DOWN” to select specific setting items,
and press “FUNC” to enter setting information, press “UP” or “DOWN” to change
parameters, press “FUNC” to confirm, and press “ESC” to exit.
5.2.6.3 Third level interface of ADVANCED
5.2.6.3.1 M.FLOW MAX
5.2.6.3.1.1 Set the flow meter's mass flow range upper limit value.
5.2.6.3.1.2 When this value is less than 100Kg/h, the K coefficient is magnified 100 times.
5.2.6.3.1.3 This value is the point of nonlinear correction 100% point.
5.2.6.3.2 M.FLOW MIN
Set the flow meter's mass flow range lower limit value.
5.2.6.3.3 V.FLOW MAX
Set the flow meter's volume flow range upper limit value.
5.2.6.3.4 V.FLOW MIN
Set the flow meter's volume flow range lower limit value.
5.2.6.3.5 FLOW K
Definition: the corresponding mass flow of Unit phase difference.
5.2.6.3.5.1 Set flow coefficient of flow meter
5.2.6.3.5.2 This parameter contains two parts: Coefficient part and index part. Flow coefficient
= coefficient x 10
index
.
5.2.6.3.5.2.1 Coefficient section setting range: 0.00000-9.99999.
5.2.6.3.5.2.2 Index part setting range: 0-4.
5.2.6.3.5.3 Matters needing attention
Flow coefficient has been set by the factory, non annual inspection or
professional personnel do not operate.
5.2.6.3.6 CAL TEMP
5.2.6.3.6.1 Process temperature of flow meter calibration.
5.2.6.3.6.2 Matters needing attention
CAL TEMP has been set by the factory, non annual inspection or professional
personnel do not operate.
5.2.6.3.7 TEMP K
t
5.2.6.3.7.1 Temperature compensation coefficient of flow meter.
5.2.6.3.7.2 Matters needing attention
The temperature coefficient K
t
has been set by the factory before the, non annual
inspection or professional personnel do not operate.
5.2.6.3.7.3 Flow temperature correction
Formula:
Q
T
= Q × (k
t
× (
t − t
0
) × 10
−5
+ 1)
Definition: Q
T
: Flow after Compensation
Q: Flow before Compensation
K
t
: Temperature correction coefficient
T: Process temperature
t
0
: Calibration temperature