background image

13

1. 

2. To measure current: Remove power from circuit, open circuit at 

measurement point, connect meter in-series in the circuit using 

the test leads, and apply power to circuit.The meter will auto-

range to display the measurement in the most appropriate range.

Do not attempt to measure more than 10A.
When measuring currents greater than 6A, a measurement

       time of 30 seconds followed by 10 minutes of recovery time

       is recommended.
CONTINUITY 

1.  Insert RED test lead into VΩ jack 

5

, and BLACK test lead 

into COM jack 

4

, and rotate function selector switch 

2

 to the 

Continuity/Resistance/Diode/Capacitance 

 setting.

NOTE:

 The meter defaults to Continuity testing in this mode. Ensure 

that the Continuity Testing icon   is visible on the display. If not, 

press the "SEL" button 

11

 until the   icon appears.

2. Remove power from circuit.
3. Test for continuity by connecting conductor or circuit with test 

leads. If resistance is measured less than 50Ω, an audible signal 

will sound and display will show a resistance value indicating 

continuity. If circuit is open, display will show "

OL

". 

OPERATING INSTRUCTIONS

DO NOT attempt to measure continuity on a live circuit.

RESISTANCE MEASUREMENTS

1.  Insert RED test lead into VΩ jack 

5

, and BLACK test lead 

into COM jack 

4

, and rotate function selector switch 

2

 to the 

Continuity/Resistance/Diode/Capacitance 

 setting.

NOTE:

 The meter defaults to Continuity testing in this mode. To 

enter Resistance testing mode, press the "SEL" button 

11

 once. 

2.  Remove power from circuit.
3.  Measure resistance by connecting test leads to circuit. The 

meter will auto-range to display the measurement in the most 

appropriate range.

Red lead

Black lead

www.calcert.com

[email protected]

1.800.544.2843

0

5

10

15

20

25

30

Содержание 092644693762

Страница 1: ...LOW IMPEDANCE AUDIBLE CONTINUITY MIN MAX RELATIVE TEMPERATURE DIODE TEST CAPACITANCE FREQUENCY Auto Ranging Digital Multimeter True RMS Measurement Technology MM720 2m IP42 CAT IV 600V CAT III 1000V 5...

Страница 2: ...3 85 7 47 7 mm Weight 12 2 oz 345 g Calibration Accurate for one year Standards Conforms to UL STD 61010 1 61010 2 030 61010 2 033 Certified to CSA STD C22 2 61010 1 61010 2 030 61010 2 033 EC EN 6101...

Страница 3: ...A 0 1 A 1 0 5 digits 6000 A 1 A 60 00mA 10 A 600 0mA 100 A 6 000A 1mA 2 0 3 digits 10 00A 0 01A 2 0 5 digits DC Current A and mA 600 0 A 0 1 A 1 0 3 digits 6000 A 1 A 60 00mA 10 A 600 0mA 100 A 6 000...

Страница 4: ...its 99 99Hz 0 01Hz 999 9Hz 0 1Hz 9 999kHz 1Hz 99 99kHz 10Hz 500 0kHz 100Hz Voltage Range 2V to 220V RMS Maximum Input 600V DC or 600V AC RMS DUTY CYCLE 1 0 to 99 9 0 1 1 2 2 digits Pulse Width 0 1 100...

Страница 5: ...uit with voltages that exceed the category based rating of this meter Do not use the meter during electrical storms or in wet weather Do not use the meter or test leads if they appear to be damaged Us...

Страница 6: ...ectrical Shock Improper use of this meter can lead to risk of electrical shock Follow all warnings and instructions detailed in this manual ENGLISH SYMBOLS ON LCD H Data Hold Audible Continuity Diode...

Страница 7: ...er 1 6000 count Reverse Contrast Display 7 RANGE button 2 Function selector switch 8 REL Relative button 3 10A jack 9 MAX MIN button 4 COM jack 10 HOLD Backlight button 5 V jack 11 SEL Select button 6...

Страница 8: ...meter in white the secondary functions are printed on the meter in orange HOLD DATA HOLD BACKLIGHT BUTTON Press HOLD 10 to hold the measurement on the display Press again to release the display to ret...

Страница 9: ...ON When the MAX MIN button 9 is pressed the meter keeps track of the minimum and maximum value of the measurement as the meter continues to take samples 1 When measuring press MAX MIN 9 and the meter...

Страница 10: ...ENT LOCATIONS Ensure the test lead shield is pressed firmly in place Failure to use the CAT III CAT IV shield increases arc flash risk TESTING IN CAT II MEASUREMENT LOCATIONS CAT III CAT IV shields ma...

Страница 11: ...TE If appears on the LCD the test leads are being applied to the circuit in reverse Swap the position of the leads to correct this NOTE When in a voltage setting and the test leads are open readings o...

Страница 12: ...rt RED test lead into 10A jack 3 and BLACK test lead into COM jack 4 and rotate function selector switch 2 to the 10A AC DC setting For mA AC DC currents 600mA Insert RED test lead into mA A jack 6 an...

Страница 13: ...utton 11 until the icon appears 2 Remove power from circuit 3 Test for continuity by connecting conductor or circuit with test leads If resistance is measured less than 50 an audible signal will sound...

Страница 14: ...otate function selector switch 2 to the Continuity Resistance Diode Capacitance setting NOTE The meter defaults to Continuity testing in this mode To enter Diode testing mode press the SEL button 11 t...

Страница 15: ...est leads across the capacitor The meter will auto range to display the measurement in the most appropriate range FREQUENCY DUTY CYCLE 1 Insert RED test lead into V jack 5 and BLACK test lead into COM...

Страница 16: ...e make contact between the thermocouple tip and the object being measured When thermocouple tip and object are in thermal equilibrium the measurement on the display will stabilize The meter will auto...

Страница 17: ...move screw from battery fuse door 2 Replace blown fuse s with A mA jack 6 800mA 1000V fast blow Klein Cat No 69390 10A jack 3 10A 1000V fast blow Klein Cat No 69034 3 Replace battery fuse door and fas...

Страница 18: ...ts mentioned in the General Specifications section allow the meter to return to normal operating conditions before using FCC IC COMPLIANCE See this product s page at for FCC compliance information Can...

Отзывы: