On pyrometers of the series KLEIBER KG 315 - K USB up to three recording modes are available. Is more
than one available, via the interface, the recording mode can be changed.

IMPORTANT:

At a change the recording mode, all measuring points recorded are deleted. Date and time must be
set again. A change is only possible if no menu is active.

6.1

Classic

In Classic mode, there are 2000 measuring values in 20 rows at 100 measuring values.

6.1.1

Choice of a measuring row or a measuring point

By pressing the SELECT

key in measuring mode it is possible to switch between the displayed storage

points. The switching order is measuring row - measuring point. By pressing

+

or

the row / point can be

selected. The SELECT

key switches to the next free storage place in the chosen row.

6.1.2

Rewriting of one or some measuring points

By pressing of the SELECT

key followed by

+

or

the row will be selected. The point to be rewritten

will be selected by SELECT

followed by

+

or

.The number of the measuring point will flicker in display.

A measuring value can be taken as described above.

The instrument will suggest the following storage point in row.

To resume at the next free storage point and finish rewriting please press SELECT .

6.2

Classic Extended

6.2.1

Configuration of the rows and values

The number of rows and values can be set freely depending on the space of the pyrometer with the help of the
software KG 315 - K USB Pyroskop Control.

IMPORTANT:

Specify the number of rows and values, all recorded measurements will be erased.

6.2.2

Advanced configuration menu

The additional parameters such as display brightness, date format, maximum temperature and so on are in
Configuration (section 5.3) at page 10 listed.

6.2.3

Date display at the measuring point display / override of the measuring points

While a previously recorded measuring point is displayed, by pressing INIT

time and date of the recording

can be displayed.

KG 315 - K USB

15

Содержание KG 315-K USB

Страница 1: ...KG 315 K USB KLEIBER Handpyrometer Operation Manual Betriebsanleitung Fast measurement fast controlling fast switching...

Страница 2: ...27 20 12 E Mail info kleiberinfrared com Internet www kleiberinfrared com 2010 2017 KLEIBER Infrared GmbH Any reproduction of this user manual or part thereof its storage on electronic media and trans...

Страница 3: ...2 2 Internal rechargeable battery charge 8 5 2 3 Changing rechargeable batteries external recharging 9 5 2 4 Rechargeable battery test 9 5 2 5 Clearing of storage 9 5 2 6 Data transmission to PC 10 5...

Страница 4: ...e rows and values 15 6 2 2 Advanced configuration menu 15 6 2 3 Date display at the measuring point display override of the measuring points 15 6 3 Ways 16 6 3 1 Basic operation on menus 16 6 3 2 Star...

Страница 5: ...erational units and accompanied by the corresponding results Listings without a sequential order are presented as bullet points and items in sub lists are preceded by dashes Safety precautions are sho...

Страница 6: ...cribed in this user manual will result in restricted warranty and liability or the loss in case of damage Damage to wearing parts e g fuses is excluded from the guarantee Warranty and liability claims...

Страница 7: ...account in the pyrometer design The safety of the pyrometer is confirmed by the decla ration of conformity and the CE mark All information related to safety is with reference to the regulations of th...

Страница 8: ...red interface IRS 315 USB with software Charging unit Voltcraft for external charging Strap with carabiner 3 2 Device models Art No Description Configuration Measuring range 08315 10010 KG 315 K USB B...

Страница 9: ...filter in holder 33152 10030 Tubular key for neutral gray filter 33152 10040 Heat protection jacket 33152 10010 Heat protection vizor 33154 10030 Wrist strap 33154 10040 Strap with carabiner 33151 100...

Страница 10: ...classic 2000 measuring values in 20 rows at 100 measuring values Data storage classic extended About 2000 measuring values in free rows and values layout Data storage ways About 1000 measuring values...

Страница 11: ...nterface Alphanumerical display Measuring value display Figure 1 User interface of KLEIBER KG 315 K USB back side KG 315 K USB www kleiberinfrared com Rubber eye piece Grey filter Ocular Strap connect...

Страница 12: ...filled up until one hundred measuring values After the row is filled the instrument starts to fill the next row from point 001 5 2 1 Recording modes The KLEIBER KG 315 K USB can be equipped with vari...

Страница 13: ...e recharged batteries must necessarily be filed with the correct polarity The recharged batteries must inserted like shown in Figure 3 Finally the battery cover is closed again and tightened the thumb...

Страница 14: ...separate manual 5 3 Configuration 5 3 1 Selection of values The configuration menu is displayed from measuring mode by pressing CONF and SELECT simultane ously In this menu the SELECT key switches to...

Страница 15: ...nly Classic Extended Ways The date format can be set in three different versions 11 Date output only Classic Extended Additional output date variant 12 Scrolling text only Ways The speed of the scroll...

Страница 16: ...pressing or in seconds 5 3 7 Automatic shutdown The instrument will shut down completely after not using it for the preset time in seconds After pressing the measuring button the instrument will maint...

Страница 17: ...ber Display 0001 CameraNo The camera number can be selected by pressing or Possible values are 1 to 1000 In mode Ways the camera number will be transferred to the PC while data transmission 5 3 11 Dis...

Страница 18: ...between 0 and 5 with 0 delivers the highest speed 5 3 15 Date and time Date and time can be selected in six steps The input is restricted to possible values 5 3 16 Factory settings Factory settings ar...

Страница 19: ...key followed by or the row will be selected The point to be rewritten will be selected by SELECT followed by or The number of the measuring point will flicker in display A measuring value can be take...

Страница 20: ...software KG 315 K USB Pyroskop Control Overlong names For most properties at the creation of Grids and Routes the names can be freely selected If these are longer than eight characters they are displ...

Страница 21: ...ossible definitions of Grids Setting the Options After selecting a Grid the Options are displayed If no Options are available this section will be skipped automatically One Options could be USEr with...

Страница 22: ...ch Route In this case this menu item is skipped First the Dynamic of the X axis is selected With DYnH Yes and DYnH No can be chosen about using Dynamic this time Grid Battery 01 Route Short Wall A Wal...

Страница 23: ...is displayed By pressing the two step button to the second step the currently measured temperature will be saved if it occurs in the selected boundaries The measured temperature will be shown until t...

Страница 24: ...or deleted Thereby are all the characteristics of a Route are defined as at the start in the same order are automatically continuously displayed Also displayed at first are the following properties S...

Страница 25: ...l e t e Y Delete Route and ProG De l e t e N Delete Route not With the button SELECT the selection is confirmed and executed Then the previous menu where it was left called back 6 3 5 Display of free...

Страница 26: ...The housing of the thermometer may be cleaned by using a non fluffing cloth with water and little detergent To clean the protective glass the lenses and the grey filter use a non fluffing cloth with e...

Страница 27: ...de Bedienung 8 5 2 1 Aufnahmemodi 8 5 2 2 Laden der Akkus im Pyrometer 8 5 2 3 Akkuwechsel Externe Akkuladung 9 5 2 4 berpr fen der Akkuspannung 9 5 2 5 L schen aller aufgenommenen Messpunkte 9 5 2 6...

Страница 28: ...n der Reihen und Werte 16 6 2 2 Erweitertes Konfigurationsmen 16 6 2 3 Datumsanzeige w hrend der Messpunktanzeige dem berschreiben von Messpunkten 17 6 3 Wege 17 6 3 1 Grundlegende Bedienung von Men s...

Страница 29: ...nd mit dem entsprechenden Ergebnis versehen Aufz hlungen ohne zeitliche Abfolge sind als Punktaufz hlungen Unteraufz hlungen als Strichaufz hlungen dargestellt Sicherheitshinweise sind mit Piktogramme...

Страница 30: ...enen bestimmungsgem en Verwendung f h ren im Schadensfall zu Einschr nkungen bzw zum Wegfall der Gew hrleistung und Haftung Sch den an Verschlei teilen z B Sicherungen sind von der Gew hrleistung ausg...

Страница 31: ...rderungen der zutreffenden Gesetze Normen und Richtlinien angewandt Die Sicherheit des Pyrometers wird durch die Konformit tserkl rung und die CE Kennzeichnung erkl rt Alle Angaben zur Sicherheit bezi...

Страница 32: ...USB mit Software Ladeger t Voltcraft f r den externen Ladevorgang Tragegurt mit Karabiner 3 2 Ger tevarianten Artikel Nr Bezeichnung Ausstattung Messbereich 08315 10010 KG 315 K USB Grundausstattung M...

Страница 33: ...ng 33152 10030 Rohrschl ssel f r Graufilter 33152 10040 W rmeschutzh lle 33152 10010 Hitzeschutzvisier 33154 10030 Handschlaufe 33154 10040 Tragegurt mit Karabiner 33151 10060 Platine mit Doppeldruckt...

Страница 34: ...ssfelddurchmesser 4 5 41 5 mm Datenspeicher Classic 2000 Messwerte in 20 Reihen zu jeweils 100 Messwerten Datenspeicher Classic Exten ded ca 2000 Messwerte in freier Reihen und Werteaufteilung Datensp...

Страница 35: ...rot Schnittstelle Alphanumerische Anzeige Messwertanzeige Abbildung 1 Bedienelemente des KLEIBER KG 315 K USB Ansicht von hinten KG 315 K USB www kleiberinfrared com Augenmuschel Graufilter Okular Bef...

Страница 36: ...enommenen Messwerte werden je nach verwendeten Modi abgespeichert Das hei t z B f r den Classicmodus dass eine Messreihe erst aufgef llt wird und danach in der n chsten Reihe mit Messwert 001 weiterge...

Страница 37: ...ndet sich auf der Unterseite des Pyrometers Die geladenen Akkus m ssen unbedingt mit der richtigen Polung eingesetzt werden Die Polung ist dem Ty penschild und der Abbildung zu entnehmen Zum Schluss w...

Страница 38: ...Daten bertragung Es ist darauf zu achten dass die abgeschr gte Seite der Schnittstelle nach oben zeigt Der USB Stecker wird mit einen freien USB Anschluss am PC verbunden Die Software KG 315 K USB Pyr...

Страница 39: ...en 8 Kameranummer nur Classic Extended Wege Das Ger t kann mit einer Kameranummer zur erleichterten Wiedererkennung versehen werden 9 Displayhelligkeit nur Classic Extended Wege Die Helligkeit Sieben...

Страница 40: ...wird bei der Bet tigung des Messtasters sofort in den Messmodus geschaltet Die f r den Energie sparmodus eingestellte Zeit muss kleiner als die Zeit f r die automatische Abschaltung sein Anzeige 0180...

Страница 41: ...nn der Messtaster in die zweite Stufe gedr ckt wird A Mittelwertmessung Die Messdaten werden kontinuierlich aufgenommen solange der Messtaster in der ersten Stufe gedr ckt wird Der st ndig aktualisier...

Страница 42: ...nem Messwert wie im Modus Classic d h nur die Uhrzeit All Ausgabe des Datums zu einem Messwert mit Uhrzeit und Datum 5 3 14 Laufschrift Anzeige 03 Scro l l Die Geschwindigkeit der Laufschrift der Dot...

Страница 43: ...maltemperatur 700 C Maximaltemperatur 1 600 C Energiesparmodus 180 s Automatische Abschaltung 3 600 s Temperatureinheit C Messmodus C Kameranummer 1 Displayhelligkeit 5 Datumsformat dnY Datumsausgabe...

Страница 44: ...der gerade ausgew hlten Reihe 6 1 2 berschreiben eines oder mehrerer Messpunkte Durch Dr cken von SELECT gefolgt von oder wird die gew nschte Messreihe ausgew hlt Danach wird mit SELECT gefolgt von od...

Страница 45: ...Ways an Zeigt das Pyrometer hingegen No Gr i d wurden auf dem Ger t noch keine Wege hinterlegt Im dem Fall sind keine Temperaturaufnahmen m glich bis Felder und Wege definiert wurden Die Definition er...

Страница 46: ...Z u g 4 Z u g 5 Grid Batt 2 Y 01 Y 02 X 0 1 X 0 2 X 0 3 X 0 4 Abbildung 5 Zwei m gliche Definitionen von Feldern Festlegen der Optionen Nach der Auswahl eines Feldes werden die Optionen angezeigt Sin...

Страница 47: ...der Erstellung der Wege f r jeden einzeln deaktiviert werden In diesen Fall wird dieser Men punkt automatisch bersprungen Zun chst wird die Dynamik der X Achse ausgew hlt Es wird mit DYnH Yes bzw DYnH...

Страница 48: ...ionen m glich Durch Dr cken des Messtasters bis zur ersten Schaltstufe wird die aktuell gemessene Temperatur ange zeigt Durch Dr cken des Messtasters bis zur zweiten Schaltstufe wird die aktuell gemes...

Страница 49: ...Wiederaufnahme L schen eines Weges Durch Dr cken der Taste INIT wird ein Men gestartet mit dem bereits fertig abgearbeitete Wege erneut gestartet beziehungsweise gel scht werden k nnen Dabei werden al...

Страница 50: ...ew hlte Aufnahme aufgerufen werden Mit den Tasten und kann zwischen ProG De l e t e Y Aufnahme l schen und ProG De l e t e N Aufnahme nicht l schen gewechselt werden Mit der Taste SELECT wird die Ausw...

Страница 51: ...en Tuch gereinigt werden Zur Reinigung der Hochtem peraturschutzscheibe der Linse im Okular und des Graufilters kann ein mit Ethylalkohol getr nktes fusselfrei es Tuch oder ein Optikputztuch verwendet...

Отзывы: