between the tip and the solder
joint.
Calibrating the Iron Temperature
The Soldering Iron should be recalibrated after changing the iron or replacing the
heating element or tip.
Method of recalibrating temperature: Use the thermometer to calibrate.
1.
Set the unit’s temperature to a certain value.
2.
When the temperature stabilizes, measure the tip’s temperature with
thermometer and write down the reading.
3.
Press “
*
” button not loose and press the “
▲
”and “
▼
” buttons
simultaneously, the soldering station enters into calibration mode.
4.
At the moment, the 100’s digit of LED display temperature is flashing. Press
the “
▲
” and “
▼
” buttons to select the value and press the “
*
” button to
select the digit. Press “
*
” button after inputting the reading. Here, the
calibration operation has been finished.
5.
When inputting calibration temperature, if the value of calibration
temperature is error and the station have protection function after input the
calibration temperature and press * button, the 100’s digit of display
temperature will flash.
6.
If the temperature still has deflection, you can repeat calibration in
accordance with above steps.
z
We recommend to use the 191/192 thermometer for measuring the tip
temperature.
z
If the unit is locked by password, it will not be able to calibrate and you
must input the right password.