Kingfisher KI6102 Series Скачать руководство пользователя страница 6

 

 

Revision 7                                         Date: 22 Jun 2020                                                                                          Page 6 of 13 

3.  Operation Instructions 

 

3.1 ON/OFF Instrument 

Press and hold [ON] to turn on instrument; the Home Display below will be 
seen on screen as shown below.  

 

 

 

 

3.2 Auto Power Off  

Auto  Power  Off  mode  is  enabled  when  instrument  is  first  turned  on.  The 
symbol, “   ” indicates that Auto Power Off mode is enabled. 

When Auto Power Off mode is enabled, instrument will turn off automatically 
if no subsequent key is pressed 10 minutes after it was turned on. 

At Home Display, press [OFF] to disable or enable Auto Power Off mode.     

 

 

 

 

The  instrument  is  now  ready  for 
measurements.  Powers  levels  of 
the respective wavelengths will be 
displayed automatically when live 
fibers are appropriately connected 
to the OLN & OLT ports. 

 
To  turn  off  instrument,  press  and 
hold [OFF]. 

 

 Indicates calibration settings are of factory default  

Indicates back light is on  

 Indicates Auto Power Off mode is enabled  

Содержание KI6102 Series

Страница 1: ...epair of the equipment This Kingfisher International product is guaranteed against defective components and workmanship for a period of 1 year from the date of delivery unless specifically stated in t...

Страница 2: ...ent 4 3 Operation Instructions 6 3 1 ON OFF Instrument 6 3 2 Auto Power Off 6 3 3 Back Light 7 3 4 Selecting Measurement Unit 7 3 5 Relative Measurement Mode 7 3 6 Threshold Setup 7 3 9 Save Retrieve...

Страница 3: ...dB 0 5 Polarization Dependent Loss dB 0 25 Linearity dB 0 1 Through Insertion Loss dB 1 5 Resolution 0 01dB Measurement unit dBm dB W Pass Fail Memory capacity 99 test Auto backlight off time 30 seco...

Страница 4: ...Revision 7 Date 22 Jun 2020 Page 4 of 13 2 The Instrument Figure 1 Front view Rubber holster Optical ports...

Страница 5: ...T Port for 1310nm upstream signal test OLT Port 1490 1550 nm downstream signal test Figure 3 Side View USB connector Connects instrument to a computer Charging Hole Insert connector of the charger to...

Страница 6: ...r Off mode is enabled instrument will turn off automatically if no subsequent key is pressed 10 minutes after it was turned on At Home Display press OFF to disable or enable Auto Power Off mode The in...

Страница 7: ...the displayed measurement unit will change to dB Press F2 to exit Relative Measurement mode 3 6 Threshold Setup 3 6 4 When a wavelength is selected press and hold OK to enter Threshold Editing mode Us...

Страница 8: ...e Display without saving Use to select the next wavelength and set the respective thresholds by repeating steps in 3 6 4 3 6 5 Press OFF to exit Threshold Editing Mode and return to Home Display The T...

Страница 9: ...ss OK to save changes or press OFF to exit Time Setting mode without saving Date format YY MM DD Time format HH MM Time Setting mode will exit automatically if no subsequent key is pressed 10 sec afte...

Страница 10: ...Test Data Browsing mode will exit automatically if no subsequent key is pressed 10 sec after it was selected 3 10 Battery When battery voltage drops below 6 7volt for Lithium battery the indicator wil...

Страница 11: ...d on surface of the sensor this will in turn result in measurement inaccuracy 4 5 Consider leaving test cords connected to the PON meter at all times This will prolong the life of the instrument ONT O...

Страница 12: ...select from the option part number listed below to order suitable AC power adapter plug if needed 6 Safety Take appropriate eye safe precautions when handling live fibre Avoid condensation The instru...

Страница 13: ...Revision 7 Date 22 Jun 2020 Page 13 of 13 Storage During prolonged storage remove batteries to eliminate the possibility of acid leakage Use only high quality batteries...