KI6680, Mini Fiber Inspection Probe User Manual
Revision 4 Date: 31/05/2018 Page 3 of 12
2.3. Specification and Standard Accessories:
Parameters
Value
Magnification
200x / 400x
Field of View
512 x 384 um
Focus method
Manual
Communication interface
WiFi802.11 / USB2.0
Software
compatible
(KI6680 Img
Viewer)
WIFI
Android 4.2 and above, iOS 9.3 & above
USB
PC Win7 and above
Power
Built-in Lithium battery rechargeable via
USB, 2.5hrs charging time (from 0 charge)
Operating/storage temp
-10 to +50 °C / -20 to + 50 °C
Relative humidity
95%
Weight
188 g including battery
Size
195(L) x 40(W) x 25(H) mm
Description
Quantity
Inspection adaptors: 1.25mm Male, 2.5mm Male, FC
Female, SC Female, LC Female
1 each
Wrist strap with captive protective cap
1
USB charger with Interchangeable International Plug
Style Adaptors
1 set
USB Cable (type A - type C)
1
Storage case for inspection adaptors
1
Operation manual
1
QA certificate
1
3. Instrument Basic Operation
3.1. Charging instrument
Connect instrument to the supplied USB charger, or to USB port of PC using the supplied USB cable. The
Battery Status Indicator
(comprises of 3
white
LEDs) show charging status as below,
3.2. Replacing instrument battery
The instrument is powered by a specific built-in Lithium battery which could be replaced as shown
below. Contact Kingfisher for replacement battery.
➊
Remove 2 screws, using a special screw driver, to
open battery door
➋
Pull battery out of battery compartment
➌
Insert new battery in an angle, and make sure that
the contacts on battery are aligned with that in battery
compartment
➍
Replace battery door and screws. Do not over-
tighten screws.
➎
Charge instrument as described in the last section
Battery status indicator pattern
Meaning
Only LED #1 lights up & blinking
Low battery. Instrument will turn off automatically if not recharged.
All 3 LEDs light up in running mode
Instrument is being charged. Max charging time needed is 2.5 hrs.
All 3 LEDs light up continually
Instrument is fully charged.
All 3 LEDs light up & blinking
Battery connection error
LED #1
contacts
Battery door
Contacts for
battery
Battery