PERFORMANCE VERIFICATION TESTS
KI2000 UM-6
-
57 -
Short Term Stability Test (optional)
1.
Connect the Light Source to the Power Meter as shown. For
better stability, ensure that connector plugged into power
meter has APC termination.
2.
Set the optical power meter to 1310 nm. 3.
On
Light
Source, enable the source and set the wavelength to 1310 nm.
4.
Let the unit warm-up for 15 minutes then note the power.
5.
Record the power every 30 seconds for 3 minutes.
6.
Calculate max-min values for stability (< 0.1dB).
7.
Record test results in Table 5.
Light Source Power Meter
Note: The same instructions apply to Short Term Stability Test on
850/1300 nm Light Sources. For improved stability, mandrel wraps
should be used in the case of 850/1300nm.
Time
Measured power,
dBm
Drift, dB
Tick max/min
values
00 sec
0.00
30 sec
60 sec
90 sec
120 sec
150 sec
180 sec