SPECIFICATIONS
KI2000 UM-6
-
42 -
Power Meter: - KI2300/KI2700 Series
Detector
Type
Response
nm
Damage
Level
dBm
Calibration
nm
Power
Range
dBm
Autotest
Sensitivity
dBm
Mid range
Linearity¹
dB
Calibration
Accuracy
2
%
Polarisation
Sensitivity
dB
Sensitivity
±30 nm dB
Total
Uncertainty
3
dB
InGaAs
600 ~ 1700
+15
780, 820, 850, 980
,
1270, 1290, 1300, 1310, 1330, 1350,
1370, 1390, 1410, 1430, 1450, 1470, 1490,
1510, 1530, 1550, 1570, 1590, 1610, 1625
,
1650
+10 ~ -60
+10 ~ -70
-40
-50
0.02
1%
(0.06 dB)
<0.005
0.03
0.3
Ge
600 ~ 1650
+25
635, 650, 660, 780, 820, 1590,
1610, 1625, 1650
850, 980, 1270, 1290, 1300,
1310, 1330, 1350, 1370, 1390,
1410, 1430, 1450, 1470, 1490,
1510, 1530, 1550, 1570
+15 ~ -50
+15 ~ -60
-40
-50
0.04
1%
(0.06 dB)
<0.005
0.03
0.5
typical
typical
typical
max
Note 1: Mid range linearity excludes top 5 dB and bottom 10 dB of range.
Note 2: Calibration condition: non coherent light, -35±5 dBm, 23±1
º
C, ±1 nm, 10±3 nm FWHM, PC ceramic connector, 100 µm fiber.
Note 3: Includes contributions of: varying optical connector types, calibration uncertainty, full temperature, dynamic range and fiber core diameter up to 200 µm.
Note 5: At calibration wavelengths in bold type
Содержание KI 2300 Series
Страница 6: ...KI2000 UM 6 4 KI2400 2800 HAND HELD FIBER SOURCE ...
Страница 7: ...KI2000 UM 6 5 KI2300 2700 HAND HELD LOSS TEST METER ...
Страница 39: ...ACCURACY CONSIDERATIONS KI2000 UM 6 37 ...
Страница 63: ...PERFORMANCE VERIFICATION TESTS KI2000 UM 6 61 Table 2 General Test Record ...
Страница 72: ...NOTES KI2000 UM 6 70 Please use these pages to write notes ...
Страница 73: ...NOTES KI2000 UM 6 71 ...
Страница 74: ...NOTES KI2000 UM 6 72 ...