The screen of measurement programming is displayed:
➢
Move the cursor on the different items and set date and time with the scrolling
knob key.
•
Beginning: date and time of the measurement beginning
•
End: date and time of the measurement end
➢
Move the cursor on the
Repeat
item to validate the value at
01
: OK appears.
➢
Press
OK
key.
The programming of measurement launching is performed.
The two last lines of the screen indicates:
•
That the memory capacity for the number of sessions is enough (
999 max
):
OK
•
That the power capacity allows the measurement:
OK
.
In case of incompatibility:
•
KO
is displayed instead of
OK:
in this case, empty the memory of the instrument.
•
Load the battery or change batteries.
If the
power adapter logo blinks
, it is imperative to use a power adapter.
When the logo stops blinking, it means that the power adapter is connected.
Note
: Using the delayed mode without programming the date and time of measurement beginning (current date and time still present)
but by programming the date and time of measurement end, the measurement starts immediately after the validation and will stop at
the programmed hour. This type of programming allows to stop the measurement and the instrument without being present.
8.4 Repetitive delayed mode – Tim.
It is possible to program the beginning and the end of the
measurement (date-time) on a daily duration (see page 37) that has
to be repeated from 2 to 10 times the following days at the same
hours. Example: following of a sound source that is marked every
days of the week.
➢
Select the icon
Tim.
at the bottom right of the screen when
set the instrument before the measurement.
➢
Press
OK
to go the programming screen.
The screen of measurement programming is displayed:
➢
Move the cursor on the different items and set date and time with the scrolling
knob key.
•
Beginning: date and time of the measurement beginning
•
End: date and time of the measurement end
➢
Move the cursor on the item
Repeat
to program the repetition value (from 2 to
10): OK appears.
➢
Press
OK
key.
The programming of measurement launching is performed.
Note:
in case of exceeding of the daily duration of programming, the instrument will correct automatically the date and time of the end
of measurement.
The two last lines of the screen indicates:
•
That the memory capacity for the number of sessions is enough (
999 max
):
OK
•
That a power adapter must be used
38
Modes of measurement launching
/
:
/
:
Mém. : OK
Alim.:
OK
Début
15
05
09
15
Fin
DIFF.-JOUR.-REPET.
15
05
16
30
1
Repet.
/
:
/
:
Mém. : OK
Alim.: OK
Début
15
05
09
15
Fin
OK
DIFF.-JOUR.-REPET.
15
05
16
30
1
Repet.
/
:
/
:
Mém. : KO
Alim.:
Début
15
05
09
15
Fin
OK
DIFF.-JOUR.-REPET.
16
05
16
30
1
Repet.
89,1
LpK
C
dB
1/2 s
.
88.7
61.4
-------
-------
dB
dB
01s
Tim.
LAeq
LCeq
dB
dB
00/00:00:00
LAeq,T
LCeq,T
E1
130
110
090
080
070
050
030
010
A C Z 63 250 1k 4k 16K
63.4
.
63
.
dB
1/8s
00/00:00:20
Hz
Tim.
1s
130
110
090
080
070
050
030
010
31.5 63 250 1k 4k 8k
---------
NR
.
T
1/8s
00/00:00:00
Tim.
/
:
/
:
Mém. : OK
Alim.:
Début
15
05
09
15
Fin
OK
DIFF.-JOUR.-REPET.
16
05
06
30
5
Repet.
/
:
/
:
Mém. : OK
Alim.:
Début
15
05
09
15
Fin
DIFF.-JOUR.-REPET.
16
05
06
30
5
Repet.