4.5 I/O
The I/O input/output allows to connect 2.5 mm jacks. It is located at the bottom of the unit behind the
rubber breastplate.
I/O 2
out:
DC continuous analogue output: 0-10V for connection to a recorder or industrial controller.
I/O 1
in
corresponds to the
input
for the launch and stop measurement from an industrial controller and is for
Leq-St – 1/1 oct – NR
measurement modes. Launching of the measurement is made for a continuous high level and stopping the measurement is made for a
continuous low level.
I/O 1 Out
corresponds to the
detection output
compared to a preset sound level and is for
L
-
Leq-St – 1/1 oct
measurement modes.
A high level appears in output when the sound peak pressure level reaches and exceeds a certain value.
For this type of output, the following items must be configured:
•
The level type: choose L, Leq or LPK depending on the mode of measurement.
•
Programming with scrolling knob key the desired level of detection by 1 dB steps.
•
Set a timer for maintaining the continuous signal detection after sub-overshoot sound level (from 1s to 10s by 1s pitch).
4.6 Autonomy
Autonomy provides information on the remaining memory
capacity and allows time measurement
according to the chosen method of
measurement.
The measurement time is calculated automatically based
on the remaining memory capacity. It is immediately
compared to the remaining capacity of the power system:
batteries, battery or AC adapter.
4.7 Instrument
This screen reminds main components of the sound level meter:
•
microphone and its serial number
•
preamplifier and its serial number
•
sound level meter housing
•
version number of the firmware.
4.8 About
This screen shows the various European standards and the accuracy class of the instrument.
It also specifies the date of the last checking and remains the next one.
Setting
14
Menu
: :
CONFIGURATION
I/O
I/O
LECTURE
PC
PROG. MESURE
INSTRUMENT
A PROPOS
AUTONOMIE
COMMENTAIRE
I/O2
Out
I/O 1
In
I/O1
Out
:
I/O
DC : 0-10V
100 dB
L/LeqA
1s
Menu
: :
CONFIGURATION
AUTONOMIE
I/O
LECTURE
PC
PROG. MESURE
INSTRUMENT
A PROPOS
AUTONOMIE
COMMENTAIRE
Memoire
Li-IOn
AUTONOMIE
87
80%
00/21:36:00
Mic.
n°:
Preamp.
n°:
SLM
n°:
S.version
AME23
12010004
PR23
12010008
DB300
12010006
E33-1-V1.0
INSTRUMENT
.