The different bars stand for:
•
LAeq
-
LCeq - LZeq:
equivalent continuous levels of sound pressure A, C and Z-weighted
•
LXeq:
equivalent continuous levels of sound pressure filtered by octave bands from 16 Hz to 16kHz (class 1) or from 31.5 Hz
to 8 kHz (class 2).
•
Measurement duration in DD/HH:MM:SS
Read values:
➢
Move the cursor in
1
with the arrows and select among
DI - global – Max - min
➢
Move the cursor in
2
with the arrows and select with scrolling knob key the value of weighting or
filter and read the result displayed in the box
3
.
A vertical line is moving on the screen and superimposes on the bar corresponding to the selected weighting
or filter.
Do the same to read the different values according to the selection in
1: DI – global – Max - min
•
DI:
equivalent continuous levels weighted and filtered for each DI logging time (01 s for example)
•
Global:
equivalent continuous levels weighted and filtered of the ongoing measurement
•
Max:
maximum equivalent continuous levels weighted and filtered reached since the
beginning of the measurement (base: DI logging time)
•
Min:
minimum equivalent continuous levels weighted and filtered reached since the
beginning of the measurement (base: DI logging time)
These different values are shown as:
•
A horizontal line for the maximum level
•
A full colored bar for the global level
•
An empty bar for the minimum level
S2:
this screen displays the results of current values that will be stored at the end of the measurement.
➢
Move the cursor in
1
with the arrows and select with the scrolling knob key among
DI - global – Max - min
Read the different values according to the selection.
•
DI:
equivalent continuous levels weighted and filtered for each DI logging time (01 s for example)
•
Global:
equivalent continuous levels weighted and filtered of the ongoing measurement
•
Max:
maximum equivalent continuous levels weighted and filtered reached since the beginning of the
measurement (base: DI logging time)
•
Min:
minimum equivalent continuous levels weighted and filtered reached since the beginning of the
measurement (base: DI logging time)
S3:
this screen remains the general format of the measurement session:
•
Date of measurement start
•
Date of measurement end: not entered because measurement is still ongoing.
•
Measurement time: not entered because measurement is still ongoing.
•
Measurement mode
1/1 oct
: Mode
•
Filters by octave bands from 16 Hz to 16kHz (class 1) or from 31.5Hz to 8 kHz (class 2).
•
LXeq logging time:
DI
•
Frequency weighting of the peak pressure level:
Lpk
•
Audio recording
: Audio
Perform some measurements
31
E1
110
090
080
070
050
030
010
A C Z 63 250 1k 4k 16K
123.4
.
250
.
dB
00/00:00:24
Hz
E1
global
1
2
3
E1
110
090
080
070
050
030
010
A C Z 63 250 1k 4k 16K
123.4
.
250
.
dB
00/00:00:24
Hz
E1
global
Max
global
min
1
LCpkmax
LAeq
LCeq
LZeq
16 Hz
31.5 Hz
63 Hz
125 Hz
250 Hz
500 Hz
1 kHz
2 kHz
4 kHz
8 kHz
16 kHz
99,5 dB
98,7 dB
31,4 dB
98,7 dB
------ dB
98,7 dB
31,4 dB
99,5 dB
98,7 dB
31,4 dB
99,5 dB
98,7 dB
31,4 dB
31,4 dB
------ dB
E2
global
1/1 oct.
31.5Hz --8kHz
1 s
C
non
Départ
Fin
Durée
Mode
F
DI
Lpk
Audio
15/05 09:15
-------
--------
15/05 13:15:45
.
E3
.