Output Terminal Insulation
Cautions when using the external voltage (Vext)
Connect the cables so that output is not shorted. See the following figures for examples of
shorted output.
A connection in which the output is shorted by the grounding of Vext (example of a prohib-
ited connection)
A connection in which the output is shorted by the shield (example of a prohibited connec-
tion)
CAUTION
The signal cable may burn out.
• Do not ground the Vext output, that is, leave it floating.
• If you are connecting the shield to the Vext side, do not connect the shield to the output ter-
minal of the PWX750ML.
+
+
–
+
–
–
2-core shielded wire
Vext
PWX
Output is shorted
by the grounding of Vext,
causing current to flow.
Ground the
positive output terminal.
Approximately the same
electric potential as
the negative output terminal
Prohibited
×
J1
+
+
–
+
–
–
2-core shielded wire
Vext
PWX
Output is shorted
by the shield,
causing current to flow.
Ground the
positive output terminal.
Prohibited
×
J1
Approximately the same
electric potential as
the negative output terminal
Содержание PWX750ML
Страница 6: ...6 PWX Appendix A Options 100 B Troubleshooting 102...
Страница 52: ...52 PWX This page is intentionally blank...
Страница 80: ...80 PWX This page is intentionally blank...
Страница 81: ...Maintenance This chapter explains how to perform cali bration DELETE...
Страница 84: ...84 PWX This page is intentionally blank...
Страница 85: ...Specifications This chapter contains the specifications and gives the dimensions of the PWX750ML...
Страница 99: ...Appendix A Options B Troubleshooting...
Страница 106: ...106 PWX This page is intentionally blank...
Страница 109: ......