Online Mode (Active Connection to N6705C Unit)
Connect to N6705C
successfully?
Can enable the
channels?
Can set the
voltage or current?
Can get waveform
after click
“Acquire”?
YES
YES
YES
YES
NO
NO
NO
NO
—
Check if N6705C mainframe has
protection information or warning
messages such as OSC/OCP/ CL+/
VL+/ OVP. Remove the protection
before test.
—
Check N6705C Settings > Protection >
Output > CV or CC.
—
Verify if there is any error message in
the display of N6705C front panel.
—
Refer to
N6705C Service Guide
for
hardware troubleshooting.
—
Verify the module set in Channel 1 is
Battery Emulator and Channel 2 is
Power Detector.
—
Check N6705C system information.
—
USB device should be auto recognized
by PC under hot plug. Open Keysight IO
Library suite version and observe the
format of the VISA address as below:
USB0::xxxxx::xxxxx::MYxxxx::0::INSTR.
—
Verify VISA address and model number
of instrument.
—
Refer to
N6705C Service Guide
when
VISA address is not populated in
Keysight IO library software.
B
Turn on the N6705C and connect to
PC using a USB cable. Launch
Keysight Power Event Monitor