Keysight N5264A Service Guide 2-5
General Product Information
Required Service Test Equipment
Required Service Test Equipment
Required Service Test Equipment (Cont’d)
Equipment
Critical Specifications
Recommended
Model or Part
Number
Alternate
Model or Part
Number
Use
a
a. P = Performance tests, A = Adjustments, T = Troubleshooting, V = System verification
Test Instruments and Software
Frequency counter
Freq: 10 MHz to 10.5 GHz
Accuracy :
0.5 ppm
53151A
Opt 001
None
P, A, T
Spectrum analyzer
Min Freq: 1 MHz
Max Freq: > 4 GHz
Resolution BW: 300 Hz
8565E
856xE
A, T
Signal generator
CW Freq: 1.185 GHz
N5181A,
Option 503
E8257D,
Option 520
P, A, T
Power meter
Accuracy: ±0.0068 dB
N1913A/14A
E4418A/B,
E4419A/B
b
b. If an accurate measurement of the dynamic accuracy specification is not required, the E4418A or E4419A can be used.
P, A, T
Power sensor
Freq: 300 kHz to 3.0 GHz
Range: –30 to +20 dBm
N8482A
8482A
P, A, T
Power sensor
Freq: 3.0 GHz to 26.5 GHz
Range: –30 to +20 dBm
N8485A
U8485A
P, A, T
Dynamic accuracy
test set
None specified
U3020AD01
None
P
Digital multi-meter
Voltage and resistance
measurement capability
Voltage resolution: 10 mV
Any
Any
T
Printer
N/A
Any printer with Microsoft Windows XP
or Windows 7 driver
P
Internal test software
c
c. The recommended model or part number for all equipment listed with a “P” in the Use column is required for proper
operation of the measurement receiver’s internal test software.
N/A
Option 897/898
None
P
Equipment
a
Critical Specifications
Recommended
Model or Part
Number
Alternate Model
or Part Number
Use
b
Calibration and Verification Kits
3.5 mm calibration kit
Freq: 10 MHz to 26.5 GHz
85052B
85052D
P,T,V