22 Installation Note N5247-90116
Description of the Upgrade
Installation Procedure for the Upgrade
-
Step 9. Install Semirigid Cables to the Noise Assembly
Refer to
for the following:
1.
Torque N5247-20058 to 10 in-lbs and add clamps 1400-1334 (x3).
2.
Torque N5247-20145 to 10 in-lbs. The other end to be connected at
14. Install the New Test Set Cables” on page 27
. Add cable ties 1400-0249
(x2).
3.
Torque N5247-20144 to 10 in-lbs. The other end to be connected at
14. Install the New Test Set Cables” on page 27
4.
Torque the 0515-2994 (x3) screws to the noise assembly and chassis to 9
in-lbs.
Figure 6
Install the Semirigid Cables to the Noise Assembly (N5247-20058,
N5247-20144, N5247-20145, 0515-2994, 1400-0249, 1400-1334)