Installation Note N5227-90118 7
Getting Prepared
Getting Prepared
To successfully install this upgrade kit, you will need the following:
— A license key - refer to
below.
— A PDF copy or a paper copy of the PNA Service Guide - refer to
“Downloading the Online PNA Service Guide”
below.
— An ESD-safe work area - refer to
“Protecting Your Workspace from
below.
“Tools Required for the Installation” on page 10
— Enough time - refer to
“About Installing the Upgrade” on page 11
.
— Test equipment for the post-upgrade adjustments and full instrument
calibration. To view the equipment list, click the Chapter 3 bookmark “Tests
and Adjustments” in the PDF Service Guide
1
.
The PNA contains extremely sensitive components that can be ruined if
mishandled. Follow instructions carefully when making cable connections,
especially wire harness connections.
The person performing the work accepts responsibility for the full cost of
the repair or replacement of damaged components.
IMPORTANT!
Before you begin this upgrade:
—
Verify your instrument’s firmware is A.13.55 or greater.
—
Verify your instrument’s IF Multiplexer (IF MUX) board, has P4, P204,
P404, P604, P804 connectors. If not, Keysight will purchase a new IF
MUX board. Refer to your instrument’s Service Guide, Chapter 7
bookmark “Removing and Replacing the A20 IF Multiplexer Board” in
the PDF Service Guide and to
“Contacting Keysight” on page 6
—
Version 6 Synthesizers: Verify your Synthesizer board is version H or
Appendix A:, “Synthesizer Board Upgrade
(N5240-60074 (with Tabs)/N5240-60076 (Without Tabs) Version F/G to
.
1. See