Installation Note N5227-90118
63
Installation Procedure for the Upgrade
Copy the license file(s) from the email(s) to the root directory of the USB
More than one license file may be copied to the USB flash drive.
3.
Insert the USB flash drive to the PNA’s USB drive slot. Within 5 seconds,
the PNA should display a small “New licenses installed” message.
Else, load the license key file(s), manually move your license file(s) to
C:\Program Files\Agilent\licensing
. It may take Keysight License
Manager an extra ~5 seconds to enable the licenses.
4.
Disconnect the USB flash drive from the PNA.
5.
On the analyzer, click or press to open the KLM software from your PNA’s
Windows taskbar by pressing
Start > More Programs > Keysight License
Manager folder > Keysight License Manager
and verify the options are
correct.
Step 32. Verify the PNA Analyzer Program is Running with the Correct
Options
1.
Start the Network Analyzer program.
2.
Once the Network Analyzer program is running:
—
Press
Help
>
About NA
and verify that Option 420 is listed in the
PNA application.
3.
After successful installation of all upgrades, some features require some
adjustments to ensure the instrument meets its specified performance.
Refer to
the Adjustments (i.e., Diagnostic Tools, Utilities, and
Adjustments) topic in the PNA Online Help:
https://rfmw.em.keysight.com/wireless/helpfiles/N52xxB/help.htm
A single license file may contain more than one feature.
Attempting to re-install a license file that is already installed may generate
a “Corrupt Media” error message. Ignore this message.
If if the option(s) have not been enabled or if your older options have not
been removed, contact Keysight Technologies. Refer to