62 Installation Note N5227-90118
Installation Procedure for the Upgrade
-
Step 30. Remove Option 419 License
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must
not
be running.
— A keyboard and mouse must be connected to the network analyzer.
Option 419 License Removal Procedure
1.
To start the Keysight License Manager, press
Start
>
Keysight License
Manager
>
Keysight License Manager
. A Keysight License Manager dialog
box will appear.
2.
Right click the on the desired option and click
Delete
.
3.
In the Keysight License Manager dialog box that appears, press or click
Yes
to confirm delete.
4.
A message displays stating that the option removal was successful.
Step 31. Enable Option 420
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must
not
be running.
— Refer to the license message you received from Keysight: Verify that the
analyzer’s model and serial numbers match those on the license message
you received from Keysight.
— A keyboard and mouse must be connected to the network analyzer.
Option Enable Procedure
1.
Locate the email(s) from Keysight which contain license file attachments.
These emails are the result of
“Step 1. Obtain a Keyword and Verify the
.
For this step, you will need a USB flash drive.
A single license file may contain more than one feature.