Installation Note
N9020-90239
49
Utilities, Adjustments, and Performance Verification Tests
Utilities, Adjustments, and Performance Verification Tests
Calibration software and specified test equipment is required to perform the adjustments and can be
used to automate the performance verification testing.
Obtain Agilent X-Series Signal Analyzer Calibration Application SW, N7814A TME Calibration
Application, version E.11.00 or later. Information on how to obtain this software can be found at:
http://www.agilent.com/find/calibrationsoftware
The following tests are required to assure the installation was performed correctly. The instrument
may not have been in spec before the upgrade was begun. Performing only these tests does not
guarantee that the analyzer meets all specifications.
Utilities Required
•
None
Adjustments Required
Performance Tests Required
Adjustment Name
10 MHz Reference Adjustment
50 MHz Calibrator Amplitude Adjustment
4800 MHz Calibrator Amplitude Adjustment
(for instruments with Option MPB)
IF Frequency Response Adjustment
Verification Test Name
Perform all performance tests