Installation Note
N9020-90239
3
Option HLB, 40 MHz to 85 MHz Analysis Bandwidth Upgrade Serial Prefix
<
MY/SG/US5233
Option HLB, 40 MHz to 85 MHz Analysis Bandwidth Upgrade
Serial Prefix
<
MY/SG/US5233
Introduction
This installation note explains how to install the hardware and provides guidelines for adjustment and
verification for Option HLB, 40 MHz to 85 MHz Analysis Bandwidth Upgrade for MXA signal
analyzers with serial number prefixes
<
MY/SG/US5233. A similar kit, Option HLC, is available for
newer MXA signal analyzers.
Installing this kit requires that the analyzer already have hardware and licenses installed to provide
analysis bandwidths up to 40 MHz. If the analyzer does not already have the necessary hardware and
licenses to provide analysis bandwidths up to 40 MHz, other upgrade kits are available.
Refer to http://www.agilent.com/find/mxa_upgrades for information of available upgrades.
If the analyzer already has Options MPB and/or DP2 installed, the upgrade to 40 MHz analysis
bandwidth, N9020AK-B40, can be installed at the same time as this kit.
There are also license-key-enabled upgrades to allow the analysis bandwidths provided by this kit to
be further extended to 125 MHz or 160 MHz. These kits are N9020AK-BU4 (85 MHz to 125 MHz
upgrade) and N9020AK-BU5 (85 MHz to 160 MHz upgrade). The N9020AK-BU4 or N9020AK-BU5
kits can also be installed at the same time as the N9020AK-HLB kit.
Products Affected:
N9020A, MXA Signal Analyzer
Serial Numbers:
<
MY52330000
<
SG52330000
<
US52330000
Options: B40,
MPB
To Be Performed By:
(X) Agilent Service Center
(X) Personnel Qualified by Agilent
( ) Customer
Estimated Installation Time:
Estimated Adjustment Time:
Estimated Verification Time:
2.0 Hours
1.5 Hours