Quick Start
Unpack and Inspect the Instrument
Samples of this product have been type tested in accordance with the Keysight
Environmental Test Manual and verified to be robust against the environmental
stresses of Storage, Transportation and End-use; those stresses include but
are not limited to temperature, humidity, shock, vibration, altitude and power
line conditions. Test Methods are aligned with IEC 60068-2 and levels are
similar to MIL-PRF-28800F Class 3.
Instrument Cooling
The following are the recommended best practices to ensure proper and safe
operating conditions:
— The primary consideration in using the chassis on a bench is ensuring
adequate ventilation for cooling. Ensure that there is at least 50 mm
(2 inches) of clearance on the sides, front, and rear of the chassis.
— For bench use, ensure that nothing is stacked on top of the instrument,
unless you use the M9484C Advanced Cooling Spacer kit. Ensure adequate
clearance is provided around all instrument vents, both air intake vents, and
air exhaust vents.
— Ensure that the fan grills are unobstructed.
— To the extent possible, install the instrument in a location with lower
ambient temperatures. For example, avoid the situation where the exhaust
air from another instrument feeds into the air intake for this instrument.
The M9484C VXG has multiple air intakes. They are located at the lower sides,
lower front and bottom of the instrument.
Содержание M9484C
Страница 4: ...4 ...
Страница 8: ...8 M9484C Startup Guide ...
Страница 20: ...20 M9484C Startup Guide Quick Start Verify Operation of the Signal Generator ...
Страница 120: ...120 M9484C Startup Guide Using Windows Tools Windows Shortcuts and Miscellaneous Tasks ...
Страница 128: ...128 M9484C Startup Guide Removable Solid State Drive SSD Instrument Security Information ...