Step 5: Verify Operation of the M9202A Module
Keysight M9202A PXIe Digitizer, Startup Guide
21
8. Verify that:
the waveform shown is stable from one acquisition to the next,
the period of the signal is 2 divisions
the pk-pk amplitude is 8 ± 0.6 divisions
SFP showing measured waveform
If a Problem is Found
1. Verify that you have set all configuration settings as shown above.
2. Verify that the RF Generator is ON and producing the desired signals at the end
of the BNC cables. This can be done with an oscilloscope.
3. If the problem occurs during Operational Verification disconnect the input signal
and note the status of the LED on the front panel; it should be GREEN. Recon-
nect the signal and select the AUTO acquisition mode; note the behavior for sub-
sequent reference.
4. Verify that the problem is reproducible.
5. Refer to
Return a Module for Service (page 10)
, concerning the details on sending
the module to Keysight for service.