background image

M9202A PXIe Wideband IF Digitizer Introduction

Follow the Startup Sequence

This Start-Up Guide is intended to lead the user through the four steps of product
installation as summarized in the diagram below. An optional fifth step shows how to
perform an operational verification of the M9202A PXIe Wideband IF Digitizer.

Step 1: Unpack and Inspect

Step 2: Verify Shipment

Step 3: Install Drivers and Software

Step 4: Install Modules

Closely follow the startup process flow in this document.
Deviating from the sequence can cause unpredictable system
behavior, damage your system, and may cause personal injury.

8

Keysight M9202A PXIe Digitizer, Startup Guide

Содержание M9202A

Страница 1: ...nd IF Digitizer 12 bit up to 2 GS s and 1 GHz bandwidth Notice This document contains references to Agilent Please note that Agilent s Test and Measurement business has become Keysight Technologies Fo...

Страница 2: ...ftware is customarily provided to the public Accordingly Keysight provides the Software to U S government customers under its standard commercial license which is embodied in its End User License Agre...

Страница 3: ...nnections are for connection to Category I sources unless otherwise marked or described in the user documentation Exercise extreme caution when a shock hazard is present Lethal voltage may be present...

Страница 4: ...A This product complies with the WEEE Directive marketing requirement The affixed product label above indicates that you must not discard this electrical electronic product in domestic household waste...

Страница 5: ...oftware 12 System Requirements 12 Hardware Requirements 13 Install the Software 13 Keysight IO Libraries Suite IOLS 13 Instrument software 13 Power up the Controller 14 Remote Controller 14 Embedded C...

Страница 6: ...Documentation Map Documentation Map 6 Keysight M9202A PXIe Digitizer Startup Guide...

Страница 7: ...ons after reviewing this information please contact your local Keysight representative or contact us through our website at www keysight com find contactus Related Documentation This Startup Guide and...

Страница 8: ...step shows how to perform an operational verification of the M9202A PXIe Wideband IF Digitizer Step 1 Unpack and Inspect Step 2 Verify Shipment Step 3 Install Drivers and Software Step 4 Install Modul...

Страница 9: ...on and conductive floor mat and heel strap combination Both types when used together provide a significant level of ESD protection Of the two only the table mat and wrist strap combination provides ad...

Страница 10: ...ioning equip ment Name and address of owner A P O box is not acceptable as a return address Product model number for example M9202A Product serial number The serial number label is located on the top...

Страница 11: ...ght MD1 High Speed Digitizer Software and Product Information DVD E2094 60003 1 Keysight IO Libraries Suite CD M9202 90001 1 Startup Guide in hard copy 5962 0476 1 Certificate of Calibration 5964 8178...

Страница 12: ...work 3 5 SP1 2 100 MB for Keysight IO Libraries Suite 100 MB Video Support for DirectX 9 graphics with 128 MB graphics memory recommended Super VGA graphics is supported Does not require graphics head...

Страница 13: ...S Keysight IO Libraries Suite IOLS which includes the Keysight Connection Expert This software is included with your shipment CD part number E2904 60003 and is also available at www keysight com find...

Страница 14: ...ou power down the chassis When you restore power power up the chassis and wait for the chassis module status indicators to be ready BEFORE you power up the PC Embedded Controller If you are using an e...

Страница 15: ...tch is in the Off Standby position 2 If the chassis has multiple fan speed settings ensure that the fans are set to auto matic Do not set the fan speed to low or turn it off 3 Position the chassis so...

Страница 16: ...necessary air circulation in the chassis 10 Plug in and power up the PXI chassis If you are using a remote controller reboot the host PC 11 Check the status LED on the module front panel It should be...

Страница 17: ...scale range is 1 V 4 dBm Frequency range 30 MHz to 1 GHz Maximum signal is 30 dBm CLK 1 IN This SMA female connector is the external clock signal input This input is AC coupled and 50 terminated The...

Страница 18: ...he instrument properties may be viewed by clicking on the desired instrument and then on the Instrument Properties tab in the right panel Agilent or Keysight Connexion Expert Interface differs dependi...

Страница 19: ...n the SFP 1 is launched it will initiate a self calibration which might take a little over 1 minute Verify that no error is shown in the lower left hand log area The image below shows the desired resu...

Страница 20: ...ure Do not exceed the maximum power level to the INPUT connector 0 5 W 1 Launch the soft front panel SFP 1 from Window Start Menu Agilent MD1 Digitizer SFP 1 11 The calibration must successfully end a...

Страница 21: ...wn above 2 Verify that the RF Generator is ON and producing the desired signals at the end of the BNC cables This can be done with an oscilloscope 3 If the problem occurs during Operational Verificati...

Страница 22: ...This information is subject to change without notice Keysight Technologies 2012 2015 Edition 2 November 2015 Printed In USA M9202 90001 www keysight com...