Installation Note E8362-90004

Step 16. Enter the New Model Number

Procedure Requirements

• The analyzer must be powered up and operating to perform this procedure.

• The Network Analyzer program must be running.

• A keyboard is required for this procedure.

• A mouse is recommended for this procedure but is not required.

Mouse Procedure

1. On the analyzer’s 

System

 menu, point to 

Service

, then click 

Option

 

Enable

.

2. Click 

Repair

.

3. Check 

Model shown is incorrect!

, then click 

Begin Repair

.

4. In the 

Model Change

 box, select the new model number, 

E8363B

.

5. Using the keyboard, enter the keyword in the box provided. The keyword is printed on the 

License Key Certificate. Enter this keyword 

exactly

 as it is printed on the certificate.

6. Click 

Enter Model

.

7. Verify that the new model number displayed is correct, then click 

YES

. If it is not correct, 

click 

NO

 and repeat the repair.

8. When the installation is complete, click 

Exit

.

Front Panel Keys Procedure

1. In the 

COMMAND

 block, press 

Menu/Dialog

.

2. In the 

NAVIGATION

 block, press the right tab and arrow keys to move over to the 

System

 

menu and down to the 

Service

 selection. Press the right tab to display the extended menu 

and the arrows to select 

Option

 

Enable

. Press 

Click

.

3. Press 

Menu/Dialog

, then press 

R

 on the keyboard to select 

Repair

.

4. Tab to 

Model shown is incorrect!

, then press 

Click

.

5. Tab to the 

Model

 

Change

 box, then use the arrows to select the new model number, 

E8363B

.

6. Tab to the 

Keyword

 box and, using the keyboard, enter the keyword. The keyword is 

printed on the License Key Certificate. Enter this keyword 

exactly

 as it is printed on the 

certificate.

7. Tab to 

Enter

 

Model

, then press 

Click

.

8. Verify that the new model number displayed is correct, then tab to 

Yes

 and press 

Click

. If 

it is not correct, tab to 

NO

, press 

Click

, and repeat the repair.

9. When the installation is complete, Tab to 

Exit

, then press 

Click

.

Содержание E8362-60109

Страница 1: ...ogies Inc All rights reserved Installation Note Extended Frequency Range Upgrade Kit For E8362B PNA Series Microwave Network Analyzers WITHOUT the Configurable Test Set Option Network Analyzer Model N...

Страница 2: ...GREEMENT WILL CONTROL Getting Assistance from Agilent By internet phone or fax get assistance with all your test and measurement needs Online assistance www agilent com find assist United States tel 1...

Страница 3: ...ed to extend the frequency range of the E8362B PNA microwave network analyzer from 20 GHz to 40 GHz Products affected E8362B without Option 014 Configurable Test Set Installation to be performed by Ag...

Страница 4: ...8 A26 Test port 2 coupler A27 Channel A first converter mixer 2 5087 7122 A30 Channel B first converter mixer N A Machine screw M3 0 x 18 for attaching A23 and A24 detectors 4 0515 0666 N A Machine sc...

Страница 5: ...ption UNL ONLY W7 A23 channel R1 detector to A28 channel R1 mixer 1 E8364 20174 W8 A24 channel R2 detector to A29 channel R2 mixer 1 E8364 20175 Parts for analyzers WITH Options UNL and 016 ONLY W7 A2...

Страница 6: ...ies within the network analyzer Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged Static discharges too small to be felt can cause...

Страница 7: ...ep 13 Replace the Front Panel Nameplate Step 14 Reinstall the Front Panel Assembly Step 15 Reinstall the Inner and Outer Covers Step 16 Enter the New Model Number Step 17 Verify that the New Model Num...

Страница 8: ...r on the certificate matches the option described by this installation note If the model serial or option number do not match the expected results you will not be able to install the option If this is...

Страница 9: ...amage the front panel assemblies 1 Disconnect the power cord if it has not already been disconnected 2 Using a T 20 TORX driver remove the strap handles item by loosening the screws item on both ends...

Страница 10: ...ation Note E8362 90004 Step 3 Remove the Inner Cover 1 Place the analyzer top side up on a flat surface 2 Using a T 10 TORX driver remove the eleven screws item 3 Lift off the cover Figure 3 Inner Cov...

Страница 11: ...om the sides of the frame CAUTION Before removing the front panel from the analyzer lift and support the front of the analyzer chassis 2 Slide the front panel over the test port connectors 3 Disconnec...

Страница 12: ...5 Disconnect the A12 Source 20 to A22 Switch Splitter Cable at the A12 Source 20 1 Using a 5 16 inch wrench disconnect the A12 source 20 to A22 switch splitter cable W1 at the A12 source 20 as shown F...

Страница 13: ...screws item that secure the receiver deck 3 Pull the latch pin item towards the opposite side of the analyzer to release the receiver deck 4 Lift the receiver deck to partially raise it then release t...

Страница 14: ...move two RF cables item from the analyzer and retain for reinstallation later 7 Remove four screws item from the mixer bracket Retain the screws for reinstallation later 8 Remove the mixer bracket wit...

Страница 15: ...Installation Note E8362 90004 15 Figure 7 Mixers Bracket Removal Figure 8 Mixers Replacement...

Страница 16: ...RF cable item from each test port coupler 3 Using a T 10 TORX driver remove eight screws items through from the test set deck front panel to release it 4 Remove the test set deck front panel from the...

Страница 17: ...e 9 2 Install a flange nut item removed from the old couplers onto each new coupler Using a 1 inch torque wrench tighten each flange nut to 72 in lbs 3 Reconnect the two cables item to the test port c...

Страница 18: ...J22 P1 BIAS T and J25 P2 BIAS T and release them from the cable clamp item 3 Disconnect the ribbon cables at the step attenuators item 4 Remove three screws item from each bracket Retain the screws f...

Страница 19: ...as Tees Replacement Reinstall the Detector Bias Tee Attenuator Brackets Refer to Figure 11 1 Place the brackets with the attenuators and bias tees attached into the analyzer as shown in Figure 11 2 Re...

Страница 20: ...ect the wrapped wire control cables from the A23 and A24 detectors A23 detector to A16 test set motherboard J5 R1 DET A24 detector to A16 test set motherboard J6 R2 DET Install the A21 SOMA 50 Refer t...

Страница 21: ...er to A23 channel R1 detector W4 A22 switch splitter to A24 channel R2 detector W37 A23 channel R1 detector to A16 test set motherboard J204 R1 W38 A24 channel R2 detector to A16 test set motherboard...

Страница 22: ...lease the receiver deck 2 Lift the receiver deck to partially lower it then release the latch pin item Lower the receiver deck to its fully lowered position and ensure that the latch pin latches in th...

Страница 23: ...A12 Source 20 to A21 SOMA 50 Cable at the A12 Source 20 1 Place the analyzer top side up on a flat surface 2 Using a 5 16 inch wrench connect the A12 source 20 to A21 SOMA 50 cable W1 at the A12 sour...

Страница 24: ...ace can be cleaned with isopropol alcohol Install the New Nameplate IMPORTANT If the Option 050 upgrade kit will be installed to extend the frequency to 50 GHz do not install the new nameplate provide...

Страница 25: ...the A3 front panel interface board 2 Slide the front panel over the test port connectors being careful to align the power switch and floppy disk drive to their corresponding front panel cutouts Ensure...

Страница 26: ...l the Inner Cover 1 Place the inner cover on the analyzer as shown There are two alignment pins on the front frame that align with holes in the front of the cover to ensure proper alignment 2 Using a...

Страница 27: ...its front panel without the handles This will damage the front panel assemblies 1 Slide the cover onto the frame 2 Using a T 20 TORX driver install the four rear panel feet item by installing the cent...

Страница 28: ...ct click NO and repeat the repair 8 When the installation is complete click Exit Front Panel Keys Procedure 1 In the COMMAND block press Menu Dialog 2 In the NAVIGATION block press the right tab and a...

Страница 29: ...OK 3 On the analyzer front panel press Preset Verify that the full frequency range of the analyzer is correct The analyzer should preset to a frequency range of 10 MHz to 40 GHz Front Panel Keys Proc...

Страница 30: ...r should now operate and phase lock over its entire frequency range If you experience difficulty with the basic functioning of the analyzer contact Agilent Refer to Getting Assistance from Agilent on...

Отзывы: