8
Installation Note E8362-90004
Equipment Required for Post-Installation Adjustments
Step 1. Verify the Model, Serial, and Option Numbers of the Analyzer
Refer to the License Key Certificate supplied with the kit:
• Verify that the analyzer’s model and serial numbers match those on the certificate.
• Verify that the option number on the certificate matches the option described by this
installation note.
If the model, serial, or option number do not match the expected results, you will not be able
to install the option. If this is the case, contact Agilent for assistance. Refer to
“Getting
Assistance from Agilent” on page 2
.
Equipment Type
Model or
Part Number
Alternate Model or
Part Number
Power meter
E4418B/E4419B
E4418A/E4419A
Power sensor, 2.4 mm
8487A
None
Adapter, 2.4 mm (f) to 2.4 mm (f)
11900B
85056-60007
RF cable, 2.4 mm (f) to 2.4 mm (f)
85133C
85133E