9 ACP Measurement
Sweep/Control
Periodic Timer Triggering:
This feature selects the internal periodic timer signal as the trigger. Trigger occurrences are set by the
Periodic Timer parameter, which is modified by the Sync Source and Offset.
The figure below shows the action of the periodic timer trigger. Before reviewing the figure, we’ll explain
some uses for the periodic trigger.
A common application is measuring periodic burst RF signals for which a trigger signal is not easily
available. For example, we might be measuring a TDMA radio which bursts every 20 ms. Let’s assume that
the 20 ms period is very consistent. Let’s also assume that we do not have an external trigger source
available that is synchronized with the period, and that the signal-to-noise ratio of the signal is not high
enough to provide a clean RF burst trigger at all of the analysis frequencies. For example, we might want to
measure spurious transmissions at an offset from the carrier that is larger than the bandwidth of the RF
burst trigger. In this application, we can set the Periodic Timer to a 20.00 ms period and adjust the offset
from that timer to position our trigger just where we want it. If we find that the 20.00 ms is not exactly right,
we can adjust the period slightly to minimize the drift between the period timer and the signal to be
measured.
A second way to use this feature would be to use Sync Source temporarily, instead of Offset. In this case,
we might tune to the signal in a narrow span and use the RF Burst trigger to synchronize the periodic timer.
Then we would turn the sync source off so that it would not miss-trigger. Miss-triggering can occur when
we are tuned so far away from the RF burst trigger that it is no longer reliable.
A third example would be to synchronize to a signal that has a reference time element of much longer
period than the period of interest. In some CDMA applications, it is useful to look at signals with a short
periodicity, by synchronizing that periodicity to the "even-second clock" edge that happens every two
seconds. Thus, we could connect the even-second clock trigger to Ext1 and use then Ext1 as the sync
source for the periodic timer.
The figure below illustrates this third example. The top trace represents the even-second clock. It causes
the periodic timer to synchronize with the leading edge shown. The analyzer trigger occurs at a time
delayed by the accumulated offset from the period trigger event. The periodic timer continues to run, and
triggers continue to occur, with a periodicity determined by the analyzer time base. The timer output
(labeled "late event") will drift away from its ideal time due to imperfect matching between the time base of
the signal being measured and the time base of the analyzer, and also because of imperfect setting of the
period parameter. But the synchronization is restored on the next even-second clock event. ("Accumulated
offset" is described in the in the Offset function section.)
946
W-CDMA & HSPA Measurement Applications Reference
Содержание E6640A
Страница 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Страница 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Страница 189: ...4 Input Output Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 189 ...
Страница 231: ...5 Mode Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 231 ...
Страница 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Страница 257: ...6 System Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 257 ...
Страница 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Страница 325: ...7 Trigger Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 325 ...
Страница 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Страница 376: ...8 Channel Power Measurement File File See File on page 258 376 W CDMA HSPA Measurement Applications Reference ...
Страница 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Страница 429: ...8 Channel Power Measurement Mode Mode See Mode on page 232 W CDMA HSPA Measurement Applications Reference 429 ...
Страница 435: ...8 Channel Power Measurement Print Print See Print on page 263 W CDMA HSPA Measurement Applications Reference 435 ...
Страница 629: ...8 Channel Power Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 629 ...
Страница 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Страница 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Страница 672: ...9 ACP Measurement File File See File on page 258 672 W CDMA HSPA Measurement Applications Reference ...
Страница 754: ...9 ACP Measurement Mode Mode See Mode on page 232 754 W CDMA HSPA Measurement Applications Reference ...
Страница 758: ...9 ACP Measurement Mode Setup Mode Setup See Mode Setup on page 251 758 W CDMA HSPA Measurement Applications Reference ...
Страница 762: ...9 ACP Measurement Print Print See Print on page 263 762 W CDMA HSPA Measurement Applications Reference ...
Страница 960: ...9 ACP Measurement System System See System on page 264 960 W CDMA HSPA Measurement Applications Reference ...
Страница 973: ...9 ACP Measurement View Display DTMB and CMMB Exciter W CDMA HSPA Measurement Applications Reference 973 ...
Страница 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Страница 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Страница 1009: ...10 Spectrum Emission Mask Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 1009 ...
Страница 1013: ...10 Spectrum Emission Mask Measurement File File See File on page 258 W CDMA HSPA Measurement Applications Reference 1013 ...
Страница 1090: ...10 Spectrum Emission Mask Measurement Mode Mode See Mode on page 232 1090 W CDMA HSPA Measurement Applications Reference ...
Страница 1096: ...10 Spectrum Emission Mask Measurement Print Print See Print on page 263 1096 W CDMA HSPA Measurement Applications Reference ...
Страница 1134: ...10 Spectrum Emission Mask Measurement Save 1134 W CDMA HSPA Measurement Applications Reference ...
Страница 1295: ...10 Spectrum Emission Mask Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 1295 ...
Страница 1332: ...10 Spectrum Emission Mask Measurement View Display 1332 W CDMA HSPA Measurement Applications Reference ...
Страница 1343: ...11 Occupied Bandwidth Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 1343 ...
Страница 1350: ...11 Occupied Bandwidth Measurement File File See File on page 258 1350 W CDMA HSPA Measurement Applications Reference ...
Страница 1398: ...11 Occupied Bandwidth Measurement Mode Mode See Mode on page 232 1398 W CDMA HSPA Measurement Applications Reference ...
Страница 1404: ...11 Occupied Bandwidth Measurement Print Print See Print on page 263 1404 W CDMA HSPA Measurement Applications Reference ...
Страница 1598: ...11 Occupied Bandwidth Measurement Sweep Control 1598 W CDMA HSPA Measurement Applications Reference ...
Страница 1599: ...11 Occupied Bandwidth Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 1599 ...
Страница 1610: ...11 Occupied Bandwidth Measurement View Display For TD SCDMA mode only 1610 W CDMA HSPA Measurement Applications Reference ...
Страница 1611: ...11 Occupied Bandwidth Measurement View Display For Bluetooth mode only W CDMA HSPA Measurement Applications Reference 1611 ...
Страница 1620: ...11 Occupied Bandwidth Measurement View Display 1620 W CDMA HSPA Measurement Applications Reference ...
Страница 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Страница 1672: ...12 Code Domain Measurement File File See File on page 258 1672 W CDMA HSPA Measurement Applications Reference ...
Страница 1762: ...12 Code Domain Measurement Mode Mode See Mode on page 232 1762 W CDMA HSPA Measurement Applications Reference ...
Страница 1766: ...12 Code Domain Measurement Mode Setup Mode Setup See Mode Setup on page 251 1766 W CDMA HSPA Measurement Applications Reference ...
Страница 1769: ...12 Code Domain Measurement Print Print See Print on page 263 W CDMA HSPA Measurement Applications Reference 1769 ...
Страница 1937: ...12 Code Domain Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 1937 ...
Страница 2019: ...13 Modulation Accuracy Composite EVM Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 2019 ...
Страница 2350: ...13 Modulation Accuracy Composite EVM Measurement View Display 2350 W CDMA HSPA Measurement Applications Reference ...
Страница 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Страница 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...
Страница 2369: ...14 QPSK EVM Measurement File File See File on page 258 W CDMA HSPA Measurement Applications Reference 2369 ...
Страница 2418: ...14 QPSK EVM Measurement Mode Mode See Mode on page 232 2418 W CDMA HSPA Measurement Applications Reference ...
Страница 2422: ...14 QPSK EVM Measurement Mode Setup Mode Setup See Mode Setup on page 251 2422 W CDMA HSPA Measurement Applications Reference ...
Страница 2426: ...14 QPSK EVM Measurement Print Print See Print on page 263 2426 W CDMA HSPA Measurement Applications Reference ...
Страница 2593: ...14 QPSK EVM Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 2593 ...
Страница 2630: ...15 Power Control Measurement File File See File on page 258 2630 W CDMA HSPA Measurement Applications Reference ...
Страница 2677: ...15 Power Control Measurement Mode Mode See Mode on page 232 W CDMA HSPA Measurement Applications Reference 2677 ...
Страница 2682: ...15 Power Control Measurement Print Print See Print on page 263 2682 W CDMA HSPA Measurement Applications Reference ...
Страница 2849: ...15 Power Control Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 2849 ...