![Keysight E6640A Скачать руководство пользователя страница 374](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_1961645374.webp)
8 Channel Power Measurement
Cont (Continuous Measurement/Sweep)
Cont (Continuous Measurement/Sweep)
Sets the analyzer for Continuous measurement operation. The single/continuous state is Meas Global so
the setting will affect all measurements. If you are Paused, pressing Cont does a Resume.
Key Path
Front-panel key
Remote Command
:INITiate:CONTinuous OFF|ON|0|1
:INITiate:CONTinuous?
Example
:INIT:CONT 0 puts analyzer in Single measurement operation.
:INIT:CONT 1 puts analyzer in Continuous measurement operation
Preset
ON
(Note that SYST:PRESet sets INIT:CONT to ON but *RST sets INIT:CONT to OFF)
State Saved
Saved in instrument state
Backwards Compatibility
Notes
For Spectrum Analysis mode in ESA and PSA, there is no Cont hardkey, instead there is a Sweep
Single/Cont key. In these analyzers, switching the Sweep Single/Cont key from Single to Cont
restarts averages (displayed average count reset to 1), but does not restart Max Hold and Min Hold.
The X-Series has Single and Cont keys in place of the SweepSingleCont key. In the X-Series, if in
single measurement, the Cont key (and INIT:CONT ON ) switches to continuous measurement, but
never restarts a measurement and never resets a sweep.
Initial S/W Revision
Prior to A.02.00
In Swept SA Measurement (Spectrum Analysis Mode):
The analyzer takes repetitive sweeps, averages, measurements, etc., when in Continuous mode. When
the average count reaches the Average/Hold Number the count stops incrementing, but the analyzer
keeps sweeping. See the Trace/Detector section for the averaging formula used both before and after the
Average/Hold Number is reached. The trigger condition must be met prior to each sweep. The type of trace
processing for multiple sweeps, is set under the Trace/Detector key, with choices of Trace Average, Max
Hold, or Min Hold.
In Other Measurements/Modes:
With Avg/Hold Num (in the Meas Setup menu) set to Off or set to On with a value of 1, a sweep is taken
after the trigger condition is met; and the analyzer continues to take new sweeps after the current sweep
has completed and the trigger condition is again met. However, with Avg/Hold Num set to On with a value
>1, multiple sweeps (data acquisitions) are taken for the measurement. The trigger condition must be met
prior to each sweep. The sweep is not stopped when the average count k equals the number N set for
Avg/Hold Num is reached, but the number k stops incrementing. A measurement average usually applies
to all traces, marker results, and numeric results. But sometimes it only applies to the numeric results.
If the analyzer is in Single measurement, pressing the Cont key does not change k and does not cause the
sweep to be reset; the only action is to put the analyzer into Continuous measurement operation.
If it is already in continuous sweep:
the INIT:CONT 1 command has no effect
374
W-CDMA & HSPA Measurement Applications Reference
Содержание E6640A
Страница 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Страница 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Страница 189: ...4 Input Output Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 189 ...
Страница 231: ...5 Mode Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 231 ...
Страница 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Страница 257: ...6 System Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 257 ...
Страница 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Страница 325: ...7 Trigger Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 325 ...
Страница 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Страница 376: ...8 Channel Power Measurement File File See File on page 258 376 W CDMA HSPA Measurement Applications Reference ...
Страница 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Страница 429: ...8 Channel Power Measurement Mode Mode See Mode on page 232 W CDMA HSPA Measurement Applications Reference 429 ...
Страница 435: ...8 Channel Power Measurement Print Print See Print on page 263 W CDMA HSPA Measurement Applications Reference 435 ...
Страница 629: ...8 Channel Power Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 629 ...
Страница 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Страница 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Страница 672: ...9 ACP Measurement File File See File on page 258 672 W CDMA HSPA Measurement Applications Reference ...
Страница 754: ...9 ACP Measurement Mode Mode See Mode on page 232 754 W CDMA HSPA Measurement Applications Reference ...
Страница 758: ...9 ACP Measurement Mode Setup Mode Setup See Mode Setup on page 251 758 W CDMA HSPA Measurement Applications Reference ...
Страница 762: ...9 ACP Measurement Print Print See Print on page 263 762 W CDMA HSPA Measurement Applications Reference ...
Страница 960: ...9 ACP Measurement System System See System on page 264 960 W CDMA HSPA Measurement Applications Reference ...
Страница 973: ...9 ACP Measurement View Display DTMB and CMMB Exciter W CDMA HSPA Measurement Applications Reference 973 ...
Страница 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Страница 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Страница 1009: ...10 Spectrum Emission Mask Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 1009 ...
Страница 1013: ...10 Spectrum Emission Mask Measurement File File See File on page 258 W CDMA HSPA Measurement Applications Reference 1013 ...
Страница 1090: ...10 Spectrum Emission Mask Measurement Mode Mode See Mode on page 232 1090 W CDMA HSPA Measurement Applications Reference ...
Страница 1096: ...10 Spectrum Emission Mask Measurement Print Print See Print on page 263 1096 W CDMA HSPA Measurement Applications Reference ...
Страница 1134: ...10 Spectrum Emission Mask Measurement Save 1134 W CDMA HSPA Measurement Applications Reference ...
Страница 1295: ...10 Spectrum Emission Mask Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 1295 ...
Страница 1332: ...10 Spectrum Emission Mask Measurement View Display 1332 W CDMA HSPA Measurement Applications Reference ...
Страница 1343: ...11 Occupied Bandwidth Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 1343 ...
Страница 1350: ...11 Occupied Bandwidth Measurement File File See File on page 258 1350 W CDMA HSPA Measurement Applications Reference ...
Страница 1398: ...11 Occupied Bandwidth Measurement Mode Mode See Mode on page 232 1398 W CDMA HSPA Measurement Applications Reference ...
Страница 1404: ...11 Occupied Bandwidth Measurement Print Print See Print on page 263 1404 W CDMA HSPA Measurement Applications Reference ...
Страница 1598: ...11 Occupied Bandwidth Measurement Sweep Control 1598 W CDMA HSPA Measurement Applications Reference ...
Страница 1599: ...11 Occupied Bandwidth Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 1599 ...
Страница 1610: ...11 Occupied Bandwidth Measurement View Display For TD SCDMA mode only 1610 W CDMA HSPA Measurement Applications Reference ...
Страница 1611: ...11 Occupied Bandwidth Measurement View Display For Bluetooth mode only W CDMA HSPA Measurement Applications Reference 1611 ...
Страница 1620: ...11 Occupied Bandwidth Measurement View Display 1620 W CDMA HSPA Measurement Applications Reference ...
Страница 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Страница 1672: ...12 Code Domain Measurement File File See File on page 258 1672 W CDMA HSPA Measurement Applications Reference ...
Страница 1762: ...12 Code Domain Measurement Mode Mode See Mode on page 232 1762 W CDMA HSPA Measurement Applications Reference ...
Страница 1766: ...12 Code Domain Measurement Mode Setup Mode Setup See Mode Setup on page 251 1766 W CDMA HSPA Measurement Applications Reference ...
Страница 1769: ...12 Code Domain Measurement Print Print See Print on page 263 W CDMA HSPA Measurement Applications Reference 1769 ...
Страница 1937: ...12 Code Domain Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 1937 ...
Страница 2019: ...13 Modulation Accuracy Composite EVM Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 2019 ...
Страница 2350: ...13 Modulation Accuracy Composite EVM Measurement View Display 2350 W CDMA HSPA Measurement Applications Reference ...
Страница 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Страница 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...
Страница 2369: ...14 QPSK EVM Measurement File File See File on page 258 W CDMA HSPA Measurement Applications Reference 2369 ...
Страница 2418: ...14 QPSK EVM Measurement Mode Mode See Mode on page 232 2418 W CDMA HSPA Measurement Applications Reference ...
Страница 2422: ...14 QPSK EVM Measurement Mode Setup Mode Setup See Mode Setup on page 251 2422 W CDMA HSPA Measurement Applications Reference ...
Страница 2426: ...14 QPSK EVM Measurement Print Print See Print on page 263 2426 W CDMA HSPA Measurement Applications Reference ...
Страница 2593: ...14 QPSK EVM Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 2593 ...
Страница 2630: ...15 Power Control Measurement File File See File on page 258 2630 W CDMA HSPA Measurement Applications Reference ...
Страница 2677: ...15 Power Control Measurement Mode Mode See Mode on page 232 W CDMA HSPA Measurement Applications Reference 2677 ...
Страница 2682: ...15 Power Control Measurement Print Print See Print on page 263 2682 W CDMA HSPA Measurement Applications Reference ...
Страница 2849: ...15 Power Control Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 2849 ...