10 Spectrum Emission Mask Measurement
FREQ Channel
When Restore Input/Output Defaults is performed, the mixer presets to the 11970A, whose Start
and Stop frequencies are 26.5 and 40 GHz respectively. The center of these two frequencies is
33.25 GHz.
Therefore, after a Restore Input/Output Defaults, if you go into External Mixing and do a Mode Preset
while in the Spectrum Analyzer Mode, the resulting Center Freq is 33.25 GHz.
State Saved
Saved in instrument state.
Min
The minimum frequency in the currently selected mixer band + 5 Hz
Max
The maximum frequency in the currently selected mixer band – 5 Hz
If the knob or step keys are being used, also depends on the value of the other three interdependent
parameters Span, Start Frequency and Stop Frequency
Initial S/W Revision
A.08.01
I/Q Center Freq
SCPI command for specifying the I/Q Center Frequency. This command will set the Center Frequency to be
used when the I/Q input is selected, even if the I/Q input is not the input which is selected at the time the
command is sent. Note that the Center Freq function in the Frequency menu on the front panel always
applies to the currently selected input.
Scope
Meas Global
Remote Command
[:SENSe]:FREQuency:IQ:CENTer <freq>
[:SENSe]:FREQuency:IQ:CENTer?
Example
FREQ:IQ:CENT: 30 MHz
Notes
This command is the same in all modes, but the parameter is Measurement Global. So the value is
independent in each mode and common across all the measurements in the mode.
Preset
0 Hz
State Saved
Saved in instrument state.
Min
–40.049995 MHz
Max
40.049995 MHz
Initial S/W Revision
Prior to A.02.00
CF Step
Changes the step size for the center frequency and start and stop frequency functions. Once a step size
has been selected and the center frequency function is active, the step keys (and the UP|DOWN
parameters for Center Frequency from remote commands) change the center frequency by the step-size
value. The step size function is useful for finding harmonics and sidebands beyond the current frequency
span of the analyzer.
Note that the start and stop frequencies also step by the CF Step value.
W-CDMA & HSPA Measurement Applications Reference
1019
Содержание E6640A
Страница 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Страница 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Страница 189: ...4 Input Output Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 189 ...
Страница 231: ...5 Mode Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 231 ...
Страница 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Страница 257: ...6 System Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 257 ...
Страница 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Страница 325: ...7 Trigger Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 325 ...
Страница 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Страница 376: ...8 Channel Power Measurement File File See File on page 258 376 W CDMA HSPA Measurement Applications Reference ...
Страница 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Страница 429: ...8 Channel Power Measurement Mode Mode See Mode on page 232 W CDMA HSPA Measurement Applications Reference 429 ...
Страница 435: ...8 Channel Power Measurement Print Print See Print on page 263 W CDMA HSPA Measurement Applications Reference 435 ...
Страница 629: ...8 Channel Power Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 629 ...
Страница 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Страница 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Страница 672: ...9 ACP Measurement File File See File on page 258 672 W CDMA HSPA Measurement Applications Reference ...
Страница 754: ...9 ACP Measurement Mode Mode See Mode on page 232 754 W CDMA HSPA Measurement Applications Reference ...
Страница 758: ...9 ACP Measurement Mode Setup Mode Setup See Mode Setup on page 251 758 W CDMA HSPA Measurement Applications Reference ...
Страница 762: ...9 ACP Measurement Print Print See Print on page 263 762 W CDMA HSPA Measurement Applications Reference ...
Страница 960: ...9 ACP Measurement System System See System on page 264 960 W CDMA HSPA Measurement Applications Reference ...
Страница 973: ...9 ACP Measurement View Display DTMB and CMMB Exciter W CDMA HSPA Measurement Applications Reference 973 ...
Страница 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Страница 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Страница 1009: ...10 Spectrum Emission Mask Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 1009 ...
Страница 1013: ...10 Spectrum Emission Mask Measurement File File See File on page 258 W CDMA HSPA Measurement Applications Reference 1013 ...
Страница 1090: ...10 Spectrum Emission Mask Measurement Mode Mode See Mode on page 232 1090 W CDMA HSPA Measurement Applications Reference ...
Страница 1096: ...10 Spectrum Emission Mask Measurement Print Print See Print on page 263 1096 W CDMA HSPA Measurement Applications Reference ...
Страница 1134: ...10 Spectrum Emission Mask Measurement Save 1134 W CDMA HSPA Measurement Applications Reference ...
Страница 1295: ...10 Spectrum Emission Mask Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 1295 ...
Страница 1332: ...10 Spectrum Emission Mask Measurement View Display 1332 W CDMA HSPA Measurement Applications Reference ...
Страница 1343: ...11 Occupied Bandwidth Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 1343 ...
Страница 1350: ...11 Occupied Bandwidth Measurement File File See File on page 258 1350 W CDMA HSPA Measurement Applications Reference ...
Страница 1398: ...11 Occupied Bandwidth Measurement Mode Mode See Mode on page 232 1398 W CDMA HSPA Measurement Applications Reference ...
Страница 1404: ...11 Occupied Bandwidth Measurement Print Print See Print on page 263 1404 W CDMA HSPA Measurement Applications Reference ...
Страница 1598: ...11 Occupied Bandwidth Measurement Sweep Control 1598 W CDMA HSPA Measurement Applications Reference ...
Страница 1599: ...11 Occupied Bandwidth Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 1599 ...
Страница 1610: ...11 Occupied Bandwidth Measurement View Display For TD SCDMA mode only 1610 W CDMA HSPA Measurement Applications Reference ...
Страница 1611: ...11 Occupied Bandwidth Measurement View Display For Bluetooth mode only W CDMA HSPA Measurement Applications Reference 1611 ...
Страница 1620: ...11 Occupied Bandwidth Measurement View Display 1620 W CDMA HSPA Measurement Applications Reference ...
Страница 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Страница 1672: ...12 Code Domain Measurement File File See File on page 258 1672 W CDMA HSPA Measurement Applications Reference ...
Страница 1762: ...12 Code Domain Measurement Mode Mode See Mode on page 232 1762 W CDMA HSPA Measurement Applications Reference ...
Страница 1766: ...12 Code Domain Measurement Mode Setup Mode Setup See Mode Setup on page 251 1766 W CDMA HSPA Measurement Applications Reference ...
Страница 1769: ...12 Code Domain Measurement Print Print See Print on page 263 W CDMA HSPA Measurement Applications Reference 1769 ...
Страница 1937: ...12 Code Domain Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 1937 ...
Страница 2019: ...13 Modulation Accuracy Composite EVM Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 2019 ...
Страница 2350: ...13 Modulation Accuracy Composite EVM Measurement View Display 2350 W CDMA HSPA Measurement Applications Reference ...
Страница 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Страница 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...
Страница 2369: ...14 QPSK EVM Measurement File File See File on page 258 W CDMA HSPA Measurement Applications Reference 2369 ...
Страница 2418: ...14 QPSK EVM Measurement Mode Mode See Mode on page 232 2418 W CDMA HSPA Measurement Applications Reference ...
Страница 2422: ...14 QPSK EVM Measurement Mode Setup Mode Setup See Mode Setup on page 251 2422 W CDMA HSPA Measurement Applications Reference ...
Страница 2426: ...14 QPSK EVM Measurement Print Print See Print on page 263 2426 W CDMA HSPA Measurement Applications Reference ...
Страница 2593: ...14 QPSK EVM Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 2593 ...
Страница 2630: ...15 Power Control Measurement File File See File on page 258 2630 W CDMA HSPA Measurement Applications Reference ...
Страница 2677: ...15 Power Control Measurement Mode Mode See Mode on page 232 W CDMA HSPA Measurement Applications Reference 2677 ...
Страница 2682: ...15 Power Control Measurement Print Print See Print on page 263 2682 W CDMA HSPA Measurement Applications Reference ...
Страница 2849: ...15 Power Control Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 2849 ...