10 Spectrum Emission Mask Measurement
Marker
:CALCulate:SEMask:MARKer[1]|2|...|12:X?
Example
CALC:SEM:MARK3:X 1.0 GHz
CALC:SEM:MARK3:X?
Notes
If no suffix is sent it will use the fundamental units for the current marker X Axis Scale. If a suffix is
sent that does not match the current marker X Axis Scale unit, an error “Invalid suffix” will be
generated.
The query returns the marker’s absolute X Axis value if the control mode is Normal. The query is
returned in the fundamental units for the current marker X Axis scale. If the marker is Off the
response is not a number.
When a Marker is turned on, it is placed center of the screen on the trace. Therefore the default value
depends on instrument condition, although the Preset/Default is defined as 1.5 GHz.
Preset
After a preset, all Markers are turned OFF, so a Marker X Axis Value query will return a not a number
(NAN).
State Saved
No
Min
–9.9E+37
Max
9.9E+37
Initial S/W Revision
Prior to A.02.00
Modified at S/W Revision
A.02.00, A.03.00
Marker X Axis Position (Remote Command Only)
Sets the marker X position in trace points. It has no effect if the control mode is Off, but is the SCPI
equivalent of entering a value if the control mode is Normal, except in trace points rather than X Axis Scale
units. The entered value is immediately translated into the current X Axis Scale units for setting the value
of the marker.
Mode
SA, WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, DTMB (CTTB), DVB-T/H, ISDB-T, CMMB,
LTE, LTETDD, Digital Cable TV,
WLAN, MSR, LTEAFDD, LTEATDD
Remote Command
:CALCulate:SEMask:MARKer[1]|2|...|12:X:POSition <real>
:CALCulate:SEMask:MARKer[1]|2|...|12:X:POSition?
Example
CALC:SEM:MARK10:X:POS 1001
CALC:SEM:MARK10:X:POS?
Notes
The query returns the marker’s absolute X Axis value in trace points if the control mode is Normal.
The value is returned as a real number, not an integer, corresponding to the translation from X Axis
Scale units to trace points . If the marker is Off the response is not a number.
When a Marker is turned on, it is placed center of the screen on the trace. Therefore the default value
depends on the instrument condition although the Preset/Default is defined as 6507 (this value
might be the expected value when all the offsets are on).
Preset
After a preset, all Markers are turned OFF, so a Marker X Axis Value query will return a not a number
(NAN).
State Saved
No
1024
W-CDMA & HSPA Measurement Applications Reference
Содержание E6640A
Страница 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Страница 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Страница 189: ...4 Input Output Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 189 ...
Страница 231: ...5 Mode Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 231 ...
Страница 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Страница 257: ...6 System Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 257 ...
Страница 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Страница 325: ...7 Trigger Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 325 ...
Страница 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Страница 376: ...8 Channel Power Measurement File File See File on page 258 376 W CDMA HSPA Measurement Applications Reference ...
Страница 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Страница 429: ...8 Channel Power Measurement Mode Mode See Mode on page 232 W CDMA HSPA Measurement Applications Reference 429 ...
Страница 435: ...8 Channel Power Measurement Print Print See Print on page 263 W CDMA HSPA Measurement Applications Reference 435 ...
Страница 629: ...8 Channel Power Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 629 ...
Страница 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Страница 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Страница 672: ...9 ACP Measurement File File See File on page 258 672 W CDMA HSPA Measurement Applications Reference ...
Страница 754: ...9 ACP Measurement Mode Mode See Mode on page 232 754 W CDMA HSPA Measurement Applications Reference ...
Страница 758: ...9 ACP Measurement Mode Setup Mode Setup See Mode Setup on page 251 758 W CDMA HSPA Measurement Applications Reference ...
Страница 762: ...9 ACP Measurement Print Print See Print on page 263 762 W CDMA HSPA Measurement Applications Reference ...
Страница 960: ...9 ACP Measurement System System See System on page 264 960 W CDMA HSPA Measurement Applications Reference ...
Страница 973: ...9 ACP Measurement View Display DTMB and CMMB Exciter W CDMA HSPA Measurement Applications Reference 973 ...
Страница 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Страница 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Страница 1009: ...10 Spectrum Emission Mask Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 1009 ...
Страница 1013: ...10 Spectrum Emission Mask Measurement File File See File on page 258 W CDMA HSPA Measurement Applications Reference 1013 ...
Страница 1090: ...10 Spectrum Emission Mask Measurement Mode Mode See Mode on page 232 1090 W CDMA HSPA Measurement Applications Reference ...
Страница 1096: ...10 Spectrum Emission Mask Measurement Print Print See Print on page 263 1096 W CDMA HSPA Measurement Applications Reference ...
Страница 1134: ...10 Spectrum Emission Mask Measurement Save 1134 W CDMA HSPA Measurement Applications Reference ...
Страница 1295: ...10 Spectrum Emission Mask Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 1295 ...
Страница 1332: ...10 Spectrum Emission Mask Measurement View Display 1332 W CDMA HSPA Measurement Applications Reference ...
Страница 1343: ...11 Occupied Bandwidth Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 1343 ...
Страница 1350: ...11 Occupied Bandwidth Measurement File File See File on page 258 1350 W CDMA HSPA Measurement Applications Reference ...
Страница 1398: ...11 Occupied Bandwidth Measurement Mode Mode See Mode on page 232 1398 W CDMA HSPA Measurement Applications Reference ...
Страница 1404: ...11 Occupied Bandwidth Measurement Print Print See Print on page 263 1404 W CDMA HSPA Measurement Applications Reference ...
Страница 1598: ...11 Occupied Bandwidth Measurement Sweep Control 1598 W CDMA HSPA Measurement Applications Reference ...
Страница 1599: ...11 Occupied Bandwidth Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 1599 ...
Страница 1610: ...11 Occupied Bandwidth Measurement View Display For TD SCDMA mode only 1610 W CDMA HSPA Measurement Applications Reference ...
Страница 1611: ...11 Occupied Bandwidth Measurement View Display For Bluetooth mode only W CDMA HSPA Measurement Applications Reference 1611 ...
Страница 1620: ...11 Occupied Bandwidth Measurement View Display 1620 W CDMA HSPA Measurement Applications Reference ...
Страница 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Страница 1672: ...12 Code Domain Measurement File File See File on page 258 1672 W CDMA HSPA Measurement Applications Reference ...
Страница 1762: ...12 Code Domain Measurement Mode Mode See Mode on page 232 1762 W CDMA HSPA Measurement Applications Reference ...
Страница 1766: ...12 Code Domain Measurement Mode Setup Mode Setup See Mode Setup on page 251 1766 W CDMA HSPA Measurement Applications Reference ...
Страница 1769: ...12 Code Domain Measurement Print Print See Print on page 263 W CDMA HSPA Measurement Applications Reference 1769 ...
Страница 1937: ...12 Code Domain Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 1937 ...
Страница 2019: ...13 Modulation Accuracy Composite EVM Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 2019 ...
Страница 2350: ...13 Modulation Accuracy Composite EVM Measurement View Display 2350 W CDMA HSPA Measurement Applications Reference ...
Страница 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Страница 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...
Страница 2369: ...14 QPSK EVM Measurement File File See File on page 258 W CDMA HSPA Measurement Applications Reference 2369 ...
Страница 2418: ...14 QPSK EVM Measurement Mode Mode See Mode on page 232 2418 W CDMA HSPA Measurement Applications Reference ...
Страница 2422: ...14 QPSK EVM Measurement Mode Setup Mode Setup See Mode Setup on page 251 2422 W CDMA HSPA Measurement Applications Reference ...
Страница 2426: ...14 QPSK EVM Measurement Print Print See Print on page 263 2426 W CDMA HSPA Measurement Applications Reference ...
Страница 2593: ...14 QPSK EVM Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 2593 ...
Страница 2630: ...15 Power Control Measurement File File See File on page 258 2630 W CDMA HSPA Measurement Applications Reference ...
Страница 2677: ...15 Power Control Measurement Mode Mode See Mode on page 232 W CDMA HSPA Measurement Applications Reference 2677 ...
Страница 2682: ...15 Power Control Measurement Print Print See Print on page 263 2682 W CDMA HSPA Measurement Applications Reference ...
Страница 2849: ...15 Power Control Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 2849 ...