10 Spectrum Emission Mask Measurement
Recall
Recall
The Recall menu lets you choose what you want to recall, and where you want to recall it from. Among the
types of files you can recall are States andTraces. In addition, an Import (Data) option lets you recall a
number of data types stored in CSV files (as used by Excel and other spreadsheet programs).
In products that run multiple instances of the X-Series Application, all instances share the same register
and file location where you want to recall from.
The default paths for Recall are data type dependent and are the same as for the Save key.
Key Path
Front-panel key
Notes
No remote command for this key specifically, but the :MMEM:LOAD command is available for
specific file types. An example is :MMEM:LOAD:STATe <filename>.
If you try to recall a State file for a mode that is not licensed or not available in the instrument, an
error message will occur and the state will not change.
Backwards Compatibility
Notes
In legacy analyzers, it was possible to load a state without affecting the trace data, limit lines or
correction data. Similarly (since User Preset is actually loading a state), it was possible to do a User
Preset without affecting the trace data, limit lines or correction data.
In the X-Series, “state” always includes all of this data; so whenever state is loaded, all of the traces,
limit lines and corrections are affected. Although this differs from previous behavior, it is desirable
behavior, and should not cause adverse issues for users.
Backwards Compatibility
Notes
Recall for the X-Series supports backward compatibility in the sense that you can recall a state file
from any X-Series model number and any version of X-Series software. This is only possible if part of
the recalling process goes through a limiting step after recalling the mode settings, at least for
settings that may vary with version number, model number, option and license differences. If you try
to recall a state file onto an instrument with less capability than what was available on the
instrument during the save, the recall will ignore the state it doesn’t support and it will limit the
recalled setting to what it allows.
Example: if the saved state includes preamp ON, but the recalling instrument does not have a
preamp; the preamp is limited to OFF. Conversely, if you save a state without a preamp, the preamp
is OFF in the state file. When this saved file is recalled on an instrument with a licensed preamp, the
preamp is changed to OFF. Another example is if the saved state has center frequency set to 20 GHz,
but the instrument recalling the saved state is a different model and only supports 13.5 GHz. In this
case, the center frequency is limited along with any other frequency based settings. Since the center
frequency can’t be preserved in this case, the recall limiting tries to at least preserve span to keep
the measurement setup as intact as possible.
It may be appropriate to issue a warning if the state is limited on the recall; warnings do not go out to
SCPI so this would only affect the manual user.
Note that there is no state file compatibility outside of the X-Series. For example, you cannot recall a
state file from ESA or PSA.
Initial S/W Revision
Prior to A.02.00
State
The Recall State menu lets you choose a register or file from which to recall the state.
W-CDMA & HSPA Measurement Applications Reference
1099
Содержание E6640A
Страница 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Страница 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Страница 189: ...4 Input Output Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 189 ...
Страница 231: ...5 Mode Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 231 ...
Страница 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Страница 257: ...6 System Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 257 ...
Страница 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Страница 325: ...7 Trigger Functions Undefined variable Primary ProductName W CDMA HSPA Measurement Applications Reference 325 ...
Страница 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Страница 376: ...8 Channel Power Measurement File File See File on page 258 376 W CDMA HSPA Measurement Applications Reference ...
Страница 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Страница 429: ...8 Channel Power Measurement Mode Mode See Mode on page 232 W CDMA HSPA Measurement Applications Reference 429 ...
Страница 435: ...8 Channel Power Measurement Print Print See Print on page 263 W CDMA HSPA Measurement Applications Reference 435 ...
Страница 629: ...8 Channel Power Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 629 ...
Страница 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Страница 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Страница 672: ...9 ACP Measurement File File See File on page 258 672 W CDMA HSPA Measurement Applications Reference ...
Страница 754: ...9 ACP Measurement Mode Mode See Mode on page 232 754 W CDMA HSPA Measurement Applications Reference ...
Страница 758: ...9 ACP Measurement Mode Setup Mode Setup See Mode Setup on page 251 758 W CDMA HSPA Measurement Applications Reference ...
Страница 762: ...9 ACP Measurement Print Print See Print on page 263 762 W CDMA HSPA Measurement Applications Reference ...
Страница 960: ...9 ACP Measurement System System See System on page 264 960 W CDMA HSPA Measurement Applications Reference ...
Страница 973: ...9 ACP Measurement View Display DTMB and CMMB Exciter W CDMA HSPA Measurement Applications Reference 973 ...
Страница 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Страница 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Страница 1009: ...10 Spectrum Emission Mask Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 1009 ...
Страница 1013: ...10 Spectrum Emission Mask Measurement File File See File on page 258 W CDMA HSPA Measurement Applications Reference 1013 ...
Страница 1090: ...10 Spectrum Emission Mask Measurement Mode Mode See Mode on page 232 1090 W CDMA HSPA Measurement Applications Reference ...
Страница 1096: ...10 Spectrum Emission Mask Measurement Print Print See Print on page 263 1096 W CDMA HSPA Measurement Applications Reference ...
Страница 1134: ...10 Spectrum Emission Mask Measurement Save 1134 W CDMA HSPA Measurement Applications Reference ...
Страница 1295: ...10 Spectrum Emission Mask Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 1295 ...
Страница 1332: ...10 Spectrum Emission Mask Measurement View Display 1332 W CDMA HSPA Measurement Applications Reference ...
Страница 1343: ...11 Occupied Bandwidth Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 1343 ...
Страница 1350: ...11 Occupied Bandwidth Measurement File File See File on page 258 1350 W CDMA HSPA Measurement Applications Reference ...
Страница 1398: ...11 Occupied Bandwidth Measurement Mode Mode See Mode on page 232 1398 W CDMA HSPA Measurement Applications Reference ...
Страница 1404: ...11 Occupied Bandwidth Measurement Print Print See Print on page 263 1404 W CDMA HSPA Measurement Applications Reference ...
Страница 1598: ...11 Occupied Bandwidth Measurement Sweep Control 1598 W CDMA HSPA Measurement Applications Reference ...
Страница 1599: ...11 Occupied Bandwidth Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 1599 ...
Страница 1610: ...11 Occupied Bandwidth Measurement View Display For TD SCDMA mode only 1610 W CDMA HSPA Measurement Applications Reference ...
Страница 1611: ...11 Occupied Bandwidth Measurement View Display For Bluetooth mode only W CDMA HSPA Measurement Applications Reference 1611 ...
Страница 1620: ...11 Occupied Bandwidth Measurement View Display 1620 W CDMA HSPA Measurement Applications Reference ...
Страница 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Страница 1672: ...12 Code Domain Measurement File File See File on page 258 1672 W CDMA HSPA Measurement Applications Reference ...
Страница 1762: ...12 Code Domain Measurement Mode Mode See Mode on page 232 1762 W CDMA HSPA Measurement Applications Reference ...
Страница 1766: ...12 Code Domain Measurement Mode Setup Mode Setup See Mode Setup on page 251 1766 W CDMA HSPA Measurement Applications Reference ...
Страница 1769: ...12 Code Domain Measurement Print Print See Print on page 263 W CDMA HSPA Measurement Applications Reference 1769 ...
Страница 1937: ...12 Code Domain Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 1937 ...
Страница 2019: ...13 Modulation Accuracy Composite EVM Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 2019 ...
Страница 2350: ...13 Modulation Accuracy Composite EVM Measurement View Display 2350 W CDMA HSPA Measurement Applications Reference ...
Страница 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Страница 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...
Страница 2369: ...14 QPSK EVM Measurement File File See File on page 258 W CDMA HSPA Measurement Applications Reference 2369 ...
Страница 2418: ...14 QPSK EVM Measurement Mode Mode See Mode on page 232 2418 W CDMA HSPA Measurement Applications Reference ...
Страница 2422: ...14 QPSK EVM Measurement Mode Setup Mode Setup See Mode Setup on page 251 2422 W CDMA HSPA Measurement Applications Reference ...
Страница 2426: ...14 QPSK EVM Measurement Print Print See Print on page 263 2426 W CDMA HSPA Measurement Applications Reference ...
Страница 2593: ...14 QPSK EVM Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 2593 ...
Страница 2630: ...15 Power Control Measurement File File See File on page 258 2630 W CDMA HSPA Measurement Applications Reference ...
Страница 2677: ...15 Power Control Measurement Mode Mode See Mode on page 232 W CDMA HSPA Measurement Applications Reference 2677 ...
Страница 2682: ...15 Power Control Measurement Print Print See Print on page 263 2682 W CDMA HSPA Measurement Applications Reference ...
Страница 2849: ...15 Power Control Measurement System System See System on page 264 W CDMA HSPA Measurement Applications Reference 2849 ...