UART/RS232 Triggering and Serial Decode
29
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
317
•
Rx 1:Data
— For use when the device under test data words are 9 bits in
length including the alert bit (the 9th bit). Triggers only when the 9th (alert)
bit is 1. The specified data byte applies to the least significant 8 bits
(excludes the 9th (alert) bit).
•
Rx 0:Data
— For use when the device under test data words are 9 bits in
length including the alert bit (the 9th bit). Triggers only when the 9th (alert)
bit is 0. The specified data byte applies to the least significant 8 bits
(excludes the 9th (alert) bit).
•
Rx X:Data
— For use when the device under test data words are 9 bits in
length including the alert bit (the 9th bit). Triggers on a data byte that you
specify regardless of the value of the 9th (alert) bit. The specified data byte
applies to the least significant 8 bits (excludes the 9th (alert) bit).
•
Similar choices are available for Tx.
•
Rx or Tx Parity Error
— Triggers on a parity error based on the parity you have
set in the UART Bus Config menu.
7
If you choose a trigger condition that includes "
Data
" in its description (for
example:
Rx Data
), then press the
Data is
softkey, and choose an equality
qualifier. You can choose equal to, not equal to, less than, or greater than a
specific data value.
8
Use the
Data
softkey to choose the data value for your trigger comparison. This
works in conjunction with the
Data is
softkey.
9
Optional: The
Burst
softkey lets you trigger on the Nth frame (1-4096) after an
idle time you select. All trigger conditions must be met for the trigger to occur.
10
If
Burst
is selected, an idle time (1 µs to 10 s) can be specified so that the
oscilloscope will look for a trigger condition only after the idle time has past.
Press the
Idle
softkey and rotate the Entry knob to set an idle time.
To display UART/RS232 serial decode, see
NOTE
If the setup does not produce a stable trigger, the UART/RS232 signals may be slow enough
that the oscilloscope is auto-triggering. Press the
[Trigger]
key, then press the
Mode
softkey
to set the trigger mode from
Auto
to
Normal
.
Содержание DSOX1202A
Страница 1: ...Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 54: ...54 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 3 Vertical Controls...
Страница 64: ...64 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 5 FFT Spectral Analysis...
Страница 76: ...76 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 6 Math Waveforms...
Страница 120: ...120 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 11 Triggers...
Страница 144: ...144 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 13 Acquisition Control...
Страница 178: ...178 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 15 Measurements...
Страница 192: ...192 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 16 Mask Testing...
Страница 196: ...196 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 17 Digital Voltmeter...
Страница 214: ...214 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 19 Waveform Generator...
Страница 226: ...226 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 20 Save Recall Setups Screens Data...
Страница 248: ...248 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 22 Utility Settings...
Страница 284: ...284 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 25 CAN Triggering and Serial Decode...
Страница 330: ...330 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide Index...