Save/Recall (Setups, Screens, Data)
20
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
217
•
Frequency Response Analysis data (*.csv)
— This creates a file of
comma-separated values for the Frequency Response Analysis results table
values. In the saved file, there are three data columns: frequency (Hz), gain
(dB), and phase (degrees). See
"To view and save the analysis results"
You can also configure the
[Quick Action]
key to save setups, screen images, or
"Configuring the [Quick Action] Key"
To save setup files
Setup files can be saved to one of 10 internal (/User Files) locations or to an
external USB storage device.
1
Press
[Save/Recall] > Save > Format
; then, turn the Entry knob to select
Setup
(*.scp)
.
2
Press the softkey in the second position and use the Entry knob to navigate to
the save location. See
"To navigate storage locations"
3
Finally, press the
Press to Save
softkey.
A message indicating whether the save was successful is displayed.
Setup files have the extension SCP. These extensions appear when using the File
Explorer (see
on page 234), but they do not appear when using the
Recall menu.
To save BMP or PNG image files
Image files can be saved to an external USB storage device.
1
Press
[Save/Recall] > Save > Format
; then, turn the Entry knob to select
8-bit
Bitmap image (*.bmp)
,
24-bit Bitmap image (*.bmp)
, or
PNG 24-bit image (*.png)
.
2
Press the softkey in the second position and use the Entry knob to navigate to
the save location. See
"To navigate storage locations"
3
Press the
Settings
softkey.
In the File Settings menu, you have these softkeys and options:
•
Setup Info
— setup information (vertical, horizontal, trigger, acquisition, math,
and display settings) is also saved in a separate file with a TXT extension.
•
Invert Grat
— the graticule in the image file has a white background instead of
the black backgound that appears on-screen.
•
Palette
— lets you choose between
Color
or
Grayscale
images.
Содержание DSOX1202A
Страница 1: ...Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 54: ...54 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 3 Vertical Controls...
Страница 64: ...64 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 5 FFT Spectral Analysis...
Страница 76: ...76 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 6 Math Waveforms...
Страница 120: ...120 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 11 Triggers...
Страница 144: ...144 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 13 Acquisition Control...
Страница 178: ...178 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 15 Measurements...
Страница 192: ...192 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 16 Mask Testing...
Страница 196: ...196 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 17 Digital Voltmeter...
Страница 214: ...214 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 19 Waveform Generator...
Страница 226: ...226 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 20 Save Recall Setups Screens Data...
Страница 248: ...248 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 22 Utility Settings...
Страница 284: ...284 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 25 CAN Triggering and Serial Decode...
Страница 330: ...330 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide Index...