104
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
11
Triggers
•
20 ns to 10 s for >< qualifier, with minimum difference of 5 ns between upper
and lower settings.
Pulse width
trigger < qualifier
time set softkey
• When the less than (
<
) qualifier is selected, the Entry knob sets the oscilloscope
to trigger on a pulse width less than the time value displayed on the softkey.
• When the time range (
><
) qualifier is selected, the Entry knob sets the upper
time range value.
Pulse width
trigger > qualifier
time set softkey
• When the greater than (
>
) qualifier is selected, the Entry knob sets the
oscilloscope to trigger on a pulse width greater than the time value displayed
on the softkey.
• When the time range (
><
) qualifier is selected, the Entry knob sets the lower
time range value.
Pattern Trigger
On DSOX1200-Series oscilloscopes, the Pattern trigger identifies a trigger
condition by looking for a specified pattern. This pattern is a logical AND
combination of the channels. Each channel can have a value of 0 (low), 1 (high), or
don't care (X). A rising or falling edge can be specified for one channel included in
the pattern.
1
Press the
[Trigger]
key.
2
In the Trigger menu, press the
Trigger Type
softkey.
3
Press the
Trigger Type
softkey; then, turn the Entry knob to select
Pattern
.
4
For each analog channel you want to include in the desired pattern, press the
Channel
softkey to select the channel.
This is the channel source for the 0, 1, X, or edge condition. As you press the
Channel
softkey (or rotate the Entry knob), the channel you select is highlighted
in the "Pattern =" line and in the upper-right corner of the display next to "Pat".
Adjust the trigger level for the selected analog channel by turning the Trigger
Level knob. The value of the trigger level is displayed in the upper-right corner
of the display.
5
For each channel you select, press the
Pattern
softkey; then, turn the Entry knob
to set the condition for that channel in the pattern.
Содержание DSOX1202A
Страница 1: ...Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 54: ...54 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 3 Vertical Controls...
Страница 64: ...64 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 5 FFT Spectral Analysis...
Страница 76: ...76 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 6 Math Waveforms...
Страница 120: ...120 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 11 Triggers...
Страница 144: ...144 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 13 Acquisition Control...
Страница 178: ...178 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 15 Measurements...
Страница 192: ...192 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 16 Mask Testing...
Страница 196: ...196 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 17 Digital Voltmeter...
Страница 214: ...214 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 19 Waveform Generator...
Страница 226: ...226 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 20 Save Recall Setups Screens Data...
Страница 248: ...248 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 22 Utility Settings...
Страница 284: ...284 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 25 CAN Triggering and Serial Decode...
Страница 330: ...330 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide Index...