Reference
24
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
265
When using segmented memory, each segment is treated as a separate
waveform. All segments for a channel are saved, then all segments of the next
(higher numbered) channel are saved. This continues until all displayed channels
are saved.
When the oscilloscope is in the Peak Detect acquisition mode, the minimum and
maximum value waveform data points are saved to the file in separate waveform
buffers. The minimum value data points are saved first; then, the maximum value
data points are saved.
BIN data - using
segmented
memory
When saving all segments, each segment has its own waveform header (see
In BIN file format, data are presented as follows:
• Channel 1 data (all segments)
• Channel 2 data (all segments)
• Channel 3 data (all segments)
• Channel 4 data (all segments)
• Digital channel data (all segments)
• Math waveform data (all segments)
When not saving all segments, the number of waveforms is equivalent to the
number of active channels (including math and digital channels, with up to seven
waveforms for each digital pod). When saving all segments, the number of
waveforms is equal to the number of active channels multiplied by the number of
segments acquired.
Binary Data in MATLAB
Binary data from an InfiniiVision oscilloscope can be imported to The MathWorks
MATLAB. You can download the appropriate MATLAB functions from the Keysight
Technologies web site at
www.keysight.com/find/1200X-Series-examples
.
Keysight provides the .m files, which need to be copied into the work directory for
MATLAB. The default work directory is C:\MATLAB7\work.
Binary Header Format
File Header
There is only one file header in a binary file. The file header consists of the
following information.
Содержание DSOX1202A
Страница 1: ...Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 54: ...54 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 3 Vertical Controls...
Страница 64: ...64 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 5 FFT Spectral Analysis...
Страница 76: ...76 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 6 Math Waveforms...
Страница 120: ...120 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 11 Triggers...
Страница 144: ...144 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 13 Acquisition Control...
Страница 178: ...178 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 15 Measurements...
Страница 192: ...192 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 16 Mask Testing...
Страница 196: ...196 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 17 Digital Voltmeter...
Страница 214: ...214 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 19 Waveform Generator...
Страница 226: ...226 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 20 Save Recall Setups Screens Data...
Страница 248: ...248 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 22 Utility Settings...
Страница 284: ...284 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 25 CAN Triggering and Serial Decode...
Страница 330: ...330 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide Index...