244
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
22
Utility Settings
1
Disconnect all inputs and waveform generator output (if available). No cables
are needed. Allow the oscilloscope to warm up before performing this
procedure.
2
Disable user calibration protection by pressing
[Utility] > Options > Auxiliary > Cal
Protect
.
3
Press
[Utility] > Service
.
4
In the Service menu, begin the user calibration by pressing the
Start User Cal
softkey.
To perform hardware self test
Pressing
[Utility] > Service > Diagnostics > Hardware Self Test
performs a series of
internal procedures to verify that the oscilloscope is operating properly.
It is recommended you run Hardware Self Test:
• After experiencing abnormal operation.
• For additional information to better describe an oscilloscope failure.
• To verify proper operation after the oscilloscope has been repaired.
Successfully passing Hardware Self Test does not guarantee 100% of the
oscilloscope's functionality. Hardware Self Test is designed to provide an 80%
confidence level that the oscilloscope is operating properly.
To perform front panel self test
Pressing
[Utility] > Service > Diagnostics > Front Panel Self Test
lets you test the front
panel keys and knobs as well as the oscilloscope display.
Follow the on-screen instructions.
To export crash log files
Pressing
[Utility] > Service > Diagnostics > Export Logs
opens the Log Export menu
where you can save crash log files to a connected USB storage device.
Crash log files are used by Keysight Technical Support when debugging
oscilloscope problems.
When there is no crash log data available to export, this softkey is ghosted
(unavailable).
Содержание DSOX1202A
Страница 1: ...Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide...
Страница 54: ...54 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 3 Vertical Controls...
Страница 64: ...64 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 5 FFT Spectral Analysis...
Страница 76: ...76 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 6 Math Waveforms...
Страница 120: ...120 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 11 Triggers...
Страница 144: ...144 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 13 Acquisition Control...
Страница 178: ...178 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 15 Measurements...
Страница 192: ...192 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 16 Mask Testing...
Страница 196: ...196 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 17 Digital Voltmeter...
Страница 214: ...214 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 19 Waveform Generator...
Страница 226: ...226 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 20 Save Recall Setups Screens Data...
Страница 248: ...248 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 22 Utility Settings...
Страница 284: ...284 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide 25 CAN Triggering and Serial Decode...
Страница 330: ...330 Keysight InfiniiVision 1200 X Series and EDUX1052A G Oscilloscopes User s Guide Index...